Instruction Manual – 5
5. Specifications
The pipettor is a high quality instrument which offers
excellent accuracy and precision. The accuracy and precision
(repeatability) of the liquid volume depends on the quality
of pipet tips used. The values for accuracy and precision
shown in the table below were obtained using manufacturer
non-filter pipet tips. Those tips are recommended for use to
ensure compatibility, accuracy and precision when pipetting.
Nominal
Volume
(μL)
Volume
(μL)
Accuracy
(%)
Precision
(%)
Non-filter
Tips (μL)
Single-channel Pipettors
2
Min.
0.1
±40.0
≤12.0
10
0.2
±12.0
≤6.0
1
±2.7
≤1.3
Max.
2
±1.5
≤0.7
10
Min.
0.5
±4.0
≤2.8
10
1.0
±2.5
≤1.8
5.0
±1.0
≤0.6
Max. 10.0
±0.5
≤0.4
20
Min.
2
±3.0
≤1.5
200
10
±1.0
≤0.5
Max.
20
±0.8
≤0.3
100
Min.
10
±1.6
≤0.80
200
50
±0.8
≤0.24
Max. 100
±0.8
≤0.20
200
Min.
20
±1.2
≤0.60
200
100
±0.8
≤0.25
Max. 200
±0.6
≤0.20
1,000
Min. 100
±1.6
≤0.40
1,000
500
±0.7
≤0.20
Max. 1,000
±0.6
≤0.15
5,000
Min. 1,000
±0.6
≤0.25
5,000
2,500
±0.6
≤0.20
Max. 5,000
±0.5
≤0.15
10,000
Min. 1,000
±2.5
≤0.6
10,000
5,000
±0.8
≤0.3
Max. 10,000
±0.5
≤0.2