Instruction Manual – 13

2. Release the pipetting pushbutton slowly and smoothly 

to aspirate the sample. Wait one second and then 

withdraw the pipet tip from the liquid.

CAUTION:

 Do not touch the used tip.

Dispensing Liquid

3. Place the end of the pipet tip against the inside wall of 

the vessel at an angle of 10° to 40°. Press the pushbutton 

smoothly to the first stop. Wait one second.

4. Press the pushbutton to the second stop to expel 

any remaining liquid. While keeping the pushbutton 

depressed, remove the pipettor from the vessel by 

drawing the pipet tip against the inside surface of 

the vessel.

5. Release the pushbutton to its starting position.
6. Eject the pipet tip by pressing the tip ejector pushbutton. 

 Remember to change the pipet tip whenever a different 

type of liquid is to be sampled.

Aspirating High-density Liquids 

When pipetting liquids of higher viscosity or lower surface 

tension than water (e.g., sera or organic solvents), a film 

of liquid may be formed on the inside of the pipet tip 

which may produce erroneous results. As the film remains 

relatively constant in successive pipetting operations with 

the same tip, this error can be eliminated by pre-rinsing 

the tip and allowing a film to form before transferring the 

first sample. This is achieved by aspirating a sample and 

dispensing it back into the same vessel. Allowing a film 

to form prior to sampling ensures optimal accuracy and 

repeatability.
This pre-rinsing operation should be repeated when 

the volume to be aspirated is changed or when a new 

pipet tip is used.

NOTE:

 Normally the degree of error resulting from viscous 

liquids is negligible if pipetting is performed slowly and 

carefully, however can be minimized further by holding the 

pipet tip in position for at least 2 seconds after aspiration 

to allow the liquid time to react to the change in pressure 

before it is dispensed. 
If the above method does not result in accurate values, 

recalibrate the pipettor in accordance with Section 9.
It is recommended to record recalibration and correction 

values, in order to facilitate reverse calibration to a standard 

liquid.

Содержание BioPette PLUS

Страница 1: ...Labnet BioPette Plus Single channel 8 channel and 12 channel Pipettors Instruction Manual This manual is available in additional languages at www labnetlink com...

Страница 2: ...mendations 4 5 Specifications 5 6 Pipettor Operation 7 7 Operating Instructions 11 8 Aspiration and Dispensing Instructions 12 9 Checking Pipetting Accuracy Parameters and Pipettor Recalibration 14 10...

Страница 3: ...0 200 L and 50 300 L Nominal Volume L Cat No Range of Pipettor Volumes L Single channel pipettors 2 P3942 2 0 1 2 10 P3942 10 0 5 10 20 P3942 20 2 20 100 P3942 100 10 100 200 P3942 200 20 200 1 000 P3...

Страница 4: ...pettor models P3942 5000 P3942 10000 3 Filters The 5 000 L and 10 000 L pipettors include a replaceable filter fitted at the bottom of the shaft which prevents aspirated liquid from entering the shaft...

Страница 5: ...me setting to be locked by pushing the ring upwards C Shaft Manufactured using high quality plastic ensuring high chemical and mechanical strength D Tip ejector For multi channel pipettors the tips ar...

Страница 6: ...in the tip Only using the pipettor in accordance with the manufacturer s instructions ensures the correct pipettor parameters are maintained After replacing the plunger or the shaft the pipettor shoul...

Страница 7: ...bility accuracy and precision when pipetting Nominal Volume L Volume L Accuracy Precision Non filter Tips L Single channel Pipettors 2 Min 0 1 40 0 12 0 10 0 2 12 0 6 0 1 2 7 1 3 Max 2 1 5 0 7 10 Min...

Страница 8: ...6 8 300 12 300 Min 50 1 6 1 5 300 150 1 2 1 0 Max 300 1 0 0 6 The accuracy and precision were obtained gravimetrically using manufacturer tips performing at least 10 measurements of distilled water a...

Страница 9: ...s approximately 1 5mm into the manifold The suspension system ensured even and airtight sealing of the pipet tips The rocking motion does not have to be performed to seal the pipet tips tightly NOTE D...

Страница 10: ...e ejector from the plastic arbor 2 Tip ejector assembly Press the tip ejector pushbutton Line the tip ejector up with the plastic arbor Turn the tip ejector bush clockwise until it has fully latched 3...

Страница 11: ...ctor 2 Tip ejector assembly Remove the tip ejector pushbutton Using a screwdriver turn the metal arbor clockwise to secure the ejector and adjust its length NOTE Leave a space of at least 1mm between...

Страница 12: ...he cap upwards to the bottom of the tip ejector M M Volume Setting 1 2 3 1 To enable volume selection set the locking ring to the lower position 2 Aspiration volume can be set by either turning the pi...

Страница 13: ...ing Instructions Observing the following recommendations will ensure maximum possible accuracy and precision of liquid sampling During operation the volume setting should be locked with black adjustme...

Страница 14: ...inse the tip several times prior to use Do not pipettor liquids with temperatures above 70 C NOTE When pipetting acids or corrosive solutions which emit vapors it is recommended to disassemble the sha...

Страница 15: ...id may be formed on the inside of the pipet tip which may produce erroneous results As the film remains relatively constant in successive pipetting operations with the same tip this error can be elimi...

Страница 16: ...tors are designed to enable recalibration and adaptation to different pipetting techniques and liquid properties e g temperature density and viscosity Periodic checks of the operation of the pipettor...

Страница 17: ...alance sensitivity should be suitable for the volume V to be measured Volume Checked V L Balance Sensitivity mg 0 1 V 10 0 001 10 V 100 0 01 100 V 1 000 0 1 V 1 000 0 1 When calculating the liquid vol...

Страница 18: ...1 94 2 06 0 63 0 0262 100 10 9 84 10 16 2 50 0 104 200 20 19 76 20 24 6 30 0 262 1 000 100 98 4 101 6 25 00 1 04 5 000 500 494 506 125 00 5 2 10 000 1 000 975 1 025 250 00 10 4 8 channel and 12 channe...

Страница 19: ...he aspirated volume Use the values given in the above table to precisely adjust the volume 4 Remove the key and replace the pipetting pushbutton by first replacing the knob then the pushbutton Determi...

Страница 20: ...than those recommended by the manufacturer check the compatibility charts and consider chemical resistance of the following plastics which form components of the pipettor PP PC POM PA PPS PVDF Sterili...

Страница 21: ...ing the instructions provided Replacement of parts should be required only occasionally and should not be required with normal pipettor use Problem Cause Solution Droplets of liquid remain in the pipe...

Страница 22: ...ts blocked The inside of the pipettor is contaminated due to aspiration of corrosive substances Unscrew the shaft set wash the parts Replace the elements with new ones if necessary Apply a small amoun...

Страница 23: ...low NOTE The replacement of the plunger requires conducting the calibration procedure according to Section 9 G L I J S C D F P G B M Item Description Model Cat No Qty Pk B Volume adjustment knob set A...

Страница 24: ...P19384 1 200 SP19385 1 1 000 SP19386 1 5 000 SP19388 1 10 000 SP19389 1 I J O ring and seal 2 SP19101 10 10 SP19102 10 20 SP19103 10 100 SP19104 10 200 SP19105 10 1 000 SP19106 10 5 000 SP19108 5 10 0...

Страница 25: ...inal material workmanship This warranty does not cover the O ring or shaft Claims for transit damage should be filed with the transportation carrier In the event this product fails within the specifie...

Страница 26: ..._________________ Date Purchased ___________________________________________ Supplier _________________________________________________ Warranty Disclaimer Unless otherwise specified all products are...

Страница 27: ...PAC I FIC Australia New Zealand t 61 427286832 Chinese Mainland t 86 21 3338 4338 f 86 21 3338 4300 India t 91 124 4604000 f 91 124 4604099 Japan t 81 3 3586 1996 f 81 3 3586 1291 Korea t 82 2 796 95...

Отзывы: