SandSET-1000 12 VDC Sludge Alarm Device  

 

                                 D15649_E-2 

Installation and Operating Instructions 

 

Copyright © 2019 Labkotec Oy 

9/13 

We reserve the right for changes without notice 

 

 

 

SET/S2, channel 1 

Poles 10 [+] ja 11 [-] 

Shortcircuit

 

20 mA 

– 24 mA

 

Sensor in the air

 

13 

– 14 mA

 

Sensor in sludge

 

12 

– 14 mA 

Sensor in the water

 

– 8 mA 

Factory setting for 
alarm point

 

approx. 11 mA 

Figure 8. Sensor current measurement 

Table 1. Sensor currents 

REPAIR AND SERVICE 

The sensor should be cleaned and the operation should also be tested 
when  emptying  the  oil  storage  chamber  or  at  least  once  every  six 
months. The easiest way to check the operation is to lift the sensor up 
in the air and to put it back to the separator. The operation is described 
in chapter 3. 

For cleaning, a mild detergent (e.g. washing-up liquid) and a scrubbing 
brush can be used. 

The mains fuse (marked 315 mAT) can be changed to another glass 
tube fuse 5 x 20 mm / 315 mAT complying IEC/EN 60127-2/3. Any other 
repair  and  service  works  on  the  device  may  be  carried  out  only  by  a 
person who has received training in Ex-i devices and is authorized by 
the manufacturer. 

In case of queries, please contact Labkotec 

Oy’s service. 

SAFETY INSTRUCTIONS 

SandSET-1000 12 VDC control unit must not be installed in potentially 
explosive atmosphere. Sensors connected to it may be installed in zone 
0, 1 or 2 potentially explosive atmospheres. 

In  case  of  installations  in  explosive  atmospheres  the  national 
requirements  and  relevant  standards  as  IEC/EN  560079-25  and/or 
IEC/EN 60079-14 must be taken into account. 

If  electrostatic  discharges  can  cause  hazards  in  the  operating 
environment,  the  device  must  be  connected  into  equipotential  ground 
according  to  requirements  with  regards  to  explosive  atmospheres. 
Equipotential grounding is made by connecting all conductive parts into 
same potential e.g. at the cable junction box. Equipotential ground must 
be earthed. 

When executing service, inspection and repair in explosive atmosphere, 
the rules in standards  IEC/EN  60079-17 and IEC/EN  60079-19 about 
instructions of Ex-devices must be obeyed. 

Содержание SandSET-1000 12 VDC

Страница 1: ...29 006 260 Fax 358 29 006 1260 28 8 2019 E mail info labkotec fi D15649_E 2 Internet www labkotec fi 1 13 Copyright 2019 Labkotec Oy We reserve the right for changes without notice SandSET 1000 12 VDC...

Страница 2: ...ntrol unit 4 2 2 SET S2 sensor 5 2 3 Cable joint 5 3 OPERATION AND SETTINGS 6 3 1 Operation 6 3 2 Factory settings 7 4 TROUBLE SHOOTING 8 5 REPAIR AND SERVICE 9 6 SAFETY INSTRUCTIONS 9 7 TECHNICAL DAT...

Страница 3: ...t SET S2 sensor junction box and installation supplies Figure 1 Sand separator alarm application SET S2 sensor is installed in the separator or tank and it gives an alarm when the sludge reaches the s...

Страница 4: ...nclosure beneath the mounting holes of the front cover The connectors of the external conductors are isolated by separating plates The plates must not be removed The plate covering the connectors must...

Страница 5: ...installation conduit equipped with a inch inner thread Variations in the water level in the well or basin must be taken into account during installation A false alarm results if the sensor is left han...

Страница 6: ...totally immersed in water Mains LED indicator is on Other LED indicators are off Relays 1 and 2 are energized Alarm The two heads of the SET S2 sensor are immersed in sludge or sand Mains LED indicato...

Страница 7: ...remains on Buzzer remains on If it has been reset earlier it will return to be on If relay 1 was already reset it will de energize again after 2 sec of continuous pressing Test will not affect relay 2...

Страница 8: ...ure 7 To do 1 Lift the sensor up in the air or immerse it into sludge and wait 10 seconds 2 Turn the SENSITIVITY potentiometer slowly anticlockwise until the sensor gives an alarm 3 Immerse the sensor...

Страница 9: ...2 3 Any other repair and service works on the device may be carried out only by a person who has received training in Ex i devices and is authorized by the manufacturer In case of queries please cont...

Страница 10: ...9 Due to non linear characteristics of the sensor voltage the interaction of both capacitance and inductance must be taken into account The table below indicates the connecting values in explosion gro...

Страница 11: ...8 DECLARATION OF COMFORMITY...

Страница 12: ...9 DECLARATION OF COMFORMITY...

Страница 13: ...10 DECLARATION OF CONFORMITY...

Отзывы: