Operation
CoaDATA 501 – Operators Manual 2.1
Page 34
2.3.1 Sample print-outs PT and calibration
General print-outs
Once a method has been selected the programmed calibration curve
parameters will be printed followed by the results. The print-out is automatic
as soon as a result has been obtained by the measuring channel.
Print-out of all parameters
A print-out of all programmed test parameters can be generated as
described in chapter 2.2.8
Utilities
.
=======================
PARAMETER-PROTOCOL
( 980 Bytes)
1-channel
V X.xx, mm/dd/yy
math-vers V XX.xx
actual date & time:
02.12.99, 14:44:00
[dd.mm.yy, hh.mm.ss]
-- device-specific: --
ser.no. j582074
Serial number of analyzer
secret no. = 11111
ntc_soll = 509
------- global: -------
Global Parameter
Analyzer name
Dealer name
parameter-ID: F1890078
Printer AUTO
computer OFF
Header OFF
Beeper ON
cuv detection ON
method 1 PT
--- method store 1 ---
Stored method parameter
PT
for PT
cuv remove detection OFF
filter.no. = 0
mode SINGLE
incubat =
60 s
start-reagent:
Lot 1
reagent =
100 ul
1st convers INTERPOLAT:
2nd convers INR
ISI = 1.05
100.0 %
11.6 s
50.0 %
17.7
s
25.0 %
29.9
s
10.0 %
66.6
s
---- method store 2 ----
aPTT