Operation
CoaDATA 501 – Operators Manual 2.1
Page 16
2.2 Method
Parameterization
2.2.1 PT-parameterization
Method parameters in the analyzer have been preset by the manufacturer.
Prior to performing clotting analysis you must update the method parameter
for the reagent used.
• Set analyzer to STANDBY mode
< 1 PT >
STANDBY
• Press
Mode
. The analyzer will request you to enter an up to 5 digit long
secret number (factory setting: 11111). For additional information refer to
chapter 2.2.8.6
Menu
secret number
.
<- secret no.:
..
_________
If the wrong number was entered STANDBY will be displayed. As soon as
the correct number has been entered, the following display will appear:
<1.conv>
Press the
arrow
key
->
to display the following menus: <
1. conv
ersion>,
<
2. conv
ersion>, <
replic
ation>, <
measur
ement>
and
<
cuv
remove
det
ec.>.
Overview PT parameterization:
<
1.conv
ersion>
<curve>
input of a 9-point calibration curve under
or
<quick>
to enter the 100%-value and slope
or
<none>
for no conversion
<
2.conv
ersion>
<INR>
input of the ISI-value for INR
or
<ratio>
input of normal value for ratio calculation
or
<none>
for no conversion
<
replic
ation>
select single
or
double determination
<single>
and the coefficient of variation
<double>
(CV 1-20%).
<
measur
ement>
start reagent volume, reagent lot. no.
and sample incubation time.
<
cuv
remove
det
ect.>
refer to chapter 3
Software
<ON>
activates the automatic cuvette detection
<OFF>
deactivates the automatic cuvette detection.
(default)