Operation
CoaDATA 501 – Operators Manual 2.1
Page 10
<-auto blanking .. keep channels clear.
The measuring channel will be adjusted for automatic cuvette detection.
(Time requirement: approximately 10 seconds).
No cuvettes must be in the measuring channels when saving the
blank values. Otherwise a wrong value is saved which might lead
to evaluation problems. Protect against external light as this might
have an impact on the blank value as well.
The method used last e.g. PT is selected.
< 1 PT >
Printer
If the printer is set to
AUTO
in the menu
UTILIT
IES the parameterization of
the selected method as well as the result of the first measurement is printed
as soon as the first measurement is completed.
Additional results will be printed automatically upon completion of a
measurement.
2.1.2 STANDBY
< 1 PT >
The selected method will be displayed.
• Press
Enter
to access the measuring mode.
• Press
Esc
to return to STANDBY.
A request for sample incubation will be displayed.
cuv
in
If there is no action the next 10 minutes, automatically the display will
change to the STANDBY mode and show the actual temperature.
37.4°C
NOTE