E
Before the treatment, switch the handpiece off and check that the needles have been
completely withdrawn into the needle cartridge. If this is not the case, dispose of the
needle cartridge immediately.
E
Ensure that needle cartridges never come into contact with contaminated items, such as
clothing for example. Contaminated needle cartridges shall not be used but rather must
be immediately disposed of.
E
During use, the handpiece, the handpiece cable and the device including the integrated
handpiece tray must be completely covered with protective film. The handpiece must be
covered with the protective film sleeve before attaching the needle cartridge (see chap-
ter "5.7 Pulling the handpiece cover on" on page 25).
E
Check at regular intervals whether the handpiece is visibly contaminated by bodily fluids
running back or by the pigment ink being used. In this case, in addition to the regular dis-
infection observe the contents of the chapters "7.4 Cleaning surfaces" on page 31
and "7.2 Material compatibilities" on page 30 as well as chapter "2.2 Product-specif-
ic safety instructions" on page 7.
E
If a local anaesthetic is being used, this should be thoroughly removed before the treat-
ment.
E
Always hold the handpiece firmly before switching it on or place it in the handpiece tray.
If an unsecured handpiece is switched on, it may move due to vibrations and cause injury
to persons or prick persons (see chapter "6.4 Switching the handpiece on and off" on
page 29).
Injuries through contaminated needles or needle cartridges can result in the transmission of
diseases (see chapter "5.6 Inserting or replacing the needle cartridge" on page 24). In
the event of being injured by a contaminated needle, seek immediate medical attention from
a physician!
In order to prevent the intensity of the treatment exceeding the extent desired:
E
Avoid risk of injury due to the needles protruding too deeply (see chapter "6.2 Setting
the needle protrusion depth" on page 28).
E
Avoid risk of injury due to the penetration frequency being too high (see chapter
"6.3 Setting the penetration frequency" on page 28).
Please be sure to inform your customer to avoid contact with recently treated skin. Recently
treated areas of skin should also be protected from
O
Contamination and
O
UV and solar radiation.
Additional irritants such as
O
Visits to swimming pools or saunas
O
Abrasive or chemical peelings
O
Hair removal at the treated areas or