Setting the needle protrusion depth
CAUTION
Risk of injury due to excessive needle protrusion
A large needle protrusion depth allows a greater penetration depth. If the needle protru-
sion depth and the penetration depth are too great, the risk of side effects, such as hae-
matoma, erythema or oedema, as well as their potential severity increases (see chap-
ter "2.4.2 Side effects" on page 11).
E
Select the needle protrusion depth on the basis of the skin characteristics of the client
and the skin area to be treated.
E
Switch the controller to operating mode.
E
Switch the handpiece on.
E
Set the needle protrusion depth by turning the moving part of the handpiece (1). A com-
plete revolution has the effect of changing the needle protrusion depth by ± 1.0 mm, de-
pending on the direction of rotation.
1
Fig. 10: Setting the needle protrusion depth
Setting the penetration frequency
CAUTION
Risk of injury due to high puncture frequency
If the puncture frequency is too high, the skin can be cut open.
E
Always start with a low puncture frequency, eg. 100 punctures per second.
E
Select the penetration frequency on the basis of the skin characteristics and suitable
for the working speed.