Chapter 2: Prerequisites
Product Service 1-800-522-7658
7
Original instructions
Proper blower selection can be determined from these exhaust requirements and
the total system static pressure loss. Contact Labconco Customer Service for
assistance in sizing a remote blower system.
Electrical Requirements
The Protector XStream Hood models feature internal wiring for the LED light
assembly and light switch. All internal wiring is terminated at the single point
wiring junction box for hook-up by a qualified electrician. The blower switch,
and light switch wires are also terminated at the single point wiring junction box
for hook-up by a qualified electrician. Refer to
Chapter 3: Getting Started
and
Appendix
C
: Protector XStream Specifications
for the wiring diagram for proper
electrical installation.
Les modèles Protector XStream Hood sont dotés d'un câblage interne pour
l'éclairage LED et l'interrupteur. Tout le câblage interne est terminé à la boîte de
jonction point de câblage unique pour le raccordement par un électricien qualifié.
Le bouton du ventilateur, et fils de l'interrupteur de lumière sont également mis
fin à la boîte de jonction point de câblage unique pour le raccordement par un
électricien qualifié. Reportez-vous au Chapitre 3: Mise en route et à l'Annexe C:
Spécifications Premier Protector pour le schéma de câblage pour l'installation
électrique correcte.
Projected Airflows and Static Pressure
Eco Energy Saving Air Foil
Flush Air Foil
Hood Type
Face
Velocity (fpm) Airflow Volumetric Rate (CFM) @
Airflow Volumetric Rate (CFM) @
Sash at
Sash at 62.5% Static Pressure (inches of water)
Static Pressure (inches of water)
Full open
open
4'
5'
6'
8'
4'
5'
6'
8'
at 28"
at 18"
Hood
Hood
Hood
Hood
Hood
Hood
Hood
Hood
XSTREAM
100
160
705, 0.26"
930, 0.32" 1150, 0.41" 1600, 0.29"
740, 0.28"
975, 0.34" 1205, 0.44" 1675, 0.31"
80
128
565, 0.17"
745, 0.20"
920, 0.26"
1280, 0.19"
590, 0.18"
780, 0.22"
965, 0.28" 1340, 0.20"
60
96
425, 0.09"
560, 0.12"
690, 0.15"
960, 0.10"
445, 0.10"
585, 0.12"
725, 0.16" 1005, 0.11"
50
80
350, 0.06"
465, 0.08"
575, 0.10"
800, 0.07"
370, 0.07"
490, 0.08"
605, 0.11"
840, 0.08"
40
64
280, 0.04"
370, 0.05"
460, 0.07"
640, 0.05"
295, 0.05"
390, 0.06"
480, 0.07"
670, 0.05"
Eco Energy Saving Air Foil
Flush Air Foil
Hood Type
Face
Velocity (fpm) Airflow Volumetric Rate (CFM) @
Airflow Volumetric Rate (CFM) @
Sash at
Sash at
Static Pressure (inches of water)
Static Pressure (inches of water)
62.50%
Full open
4'
5'
6'
8'
4'
5'
6'
8'
open at 18"
at 28"
Hood
Hood
Hood
Hood
Hood
Hood
Hood
Hood
XSTREAM
100
62
440, 0.10"
580, 0.12"
720, 0.16"
1000, 0.11"
475, 0.11"
625, 0.13"
775, 0.17" 1075, 0.12"
80
50
350, 0.06"
465, 0.08"
575, 0.10"
800, 0.07"
380, 0.07"
500, 0.08"
620, 0.11"
860, 0.08"
60
38
265, 0.04"
350, 0.05"
430, 0.06"
600, 0.04"
285, 0.04"
375, 0.05"
465, 0.06"
645, 0.04"