Chapter 8: Troubleshooting
Product Service 1-800-522-7658
37
Original instructions
PROBLEM
CAUSE
CORRECTIVE ACTION
Cable is frayed or has
slipped off the
pulleys.
Re-install, cable must be replaced
immediately if damaged.
Electrical duplex
outlets no longer
have power
Wires not connected
or faulty duplex.
Check wire connection or replace
duplex.
Circuit breakers
tripped in building
electrical supply.
Reset circuit breakers.
Service valves no
longer operate
Faulty building
supply.
Inspect building supply shut off valves
and appropriate pressures below the
recommended 40 PSI.
Valve no longer
operates.
Replace valve and check for leaks.
Supply line or outlet
line has leaks.
Inspect line for leaks and fix any
leaking plumbing connections.
Audible Alarm
Sounding
Both sashes are too
high.
Close one sash completely.