Summary of Operation 

The Labconco extractor reduces extraction time from 16 hours to about four 
hours through its efficient refluxing system.  The operation is carried out by 
using a Pyrex

®

 beaker as the solvent chamber.  The procedure involves 

placing the samples between the boiling solvent and a cold surface.  As 
boiling continues, the solvent vaporizes, condenses on the cold surface and 
washes back through the samples to the boiling solvent below. 
 

Procedures 

The following section provides a general outline of steps necessary to operate 
a routine extraction on the Labconco Goldfisch Fat Extractor.  Actual test 
methods, sample preparation, solvent choices, etc. are determined by official 
methods of the A.O.A.C. and should be referred to in all cases.

1

 

 
(1)

 

Turn on cold water supply.  Check visual flow at the drain area.  
Water flow will be adjusted later after system is operating to obtain 
normal reflux rate.  Heater and water flow rates may be adjusted to 
vary reflux rates according to the operator’s preference. 

 
(2)

 

Prepare sample(s) per standard method for the product under 
examination.  See assembly drawing in Appendix A. 

 
(3)

 

Place required quantity of sample in the extraction thimbles.  Filter 
paper or fritted glass thimbles may be substituted according to 
operator preference. 

 
(4)

 

Place thimble containing sample into Pyrex sample tube. 

 
(5)

 

Insert sample tube assembly into clip from the front.  Push upward 
until bulb of sample container meets holding clip. 

 

                                                 

1

 For more information, see “Official Methods of Analysis” of the Association of Official 

Analytical Chemists. 

Содержание 3500100

Страница 1: ...Goldfisch Fat Extraction Apparatus Models 3500100...

Страница 2: ...30 a m and 5 30 p m Central Standard Time Part 3518100 Rev L ECO L123 Warranty Labconco Corporation provides a warranty to the original buyer for the repair or replacement of parts and reasonable labo...

Страница 3: ...n 2 CHAPTER 2 INSTALLATION 4 Preparation 4 Location 4 Utility Connections 4 Initial Start Up and Check Out 4 Gasket Assembly 5 CHAPTER 3 NORMAL OPERATION ROUTINE MAINTENANCE 6 Summary of Operation 6 P...

Страница 4: ...have been checked against the following component list and the equipment has been installed and tested As shipped the carton should contain the following Part Number Description 35001 Goldfisch Fat Ex...

Страница 5: ...er to the condenser heads Water exits through the water control valve located on the upper left hand portion of the unit And flows through the manifold assembly depositing cool water in the condenser...

Страница 6: ...Chapter 1 Introduction Product Service 1 800 522 7658 3 Figure 1...

Страница 7: ...ood with proper airflow to handle the extremely flammable solvents that are used during the Goldfisch Extractor s normal operation Utility Connections 1 Once the unit has been placed in its final loca...

Страница 8: ...ing parts package which should be taped to the unit Assemble the gasket wave washer retaining ring and glassware according to assembly drawing instructions as shown in Appendix A Hand tighten the reta...

Страница 9: ...tor Actual test methods sample preparation solvent choices etc are determined by official methods of the A O A C and should be referred to in all cases 1 1 Turn on cold water supply Check visual flow...

Страница 10: ...witch should be loaded with a beaker of solvent or water to load the heater and increase its service life 10 Pre heating the heaters for about 8 to 10 minutes prior to starting samples will reduce wai...

Страница 11: ...h the loss reported as the ether extract Official methods should be referred to for specific test procedure of the indirect method in all cases Maintenance All of the parts contained in the operating...

Страница 12: ...5163 1 Switch infinite control 35046 Pkg of 6 Beaker glass Pyrex tall form 5 high 35053 Pkg of 6 Beaker glass Pyrex 100 cc 3 1 4 high 35054 Ring beaker 35059 Pkg of 6 Extraction thimble 22 x 80 mm 350...

Страница 13: ...Appendix A Replacement Parts Product Service 1 800 522 7658 10...

Страница 14: ...Product Service 1 800 522 7658 11 A Ap pp pe en nd di ix x B B D Di im me en ns si io on ns s 35 7 26 9 9...

Страница 15: ...No 35001 operates on 115 volts 50 60 Hz max at 5 2 amps It is shipped complete with power cord and grounded 3 wire plug Standard features are as follows 1 Main power on off 2 Full range variable heat...

Страница 16: ...Product Service 1 800 522 7658 13 A Ap pp pe en nd di ix x D D W Wi ir ri in ng g D Di ia ag gr ra am m...

Отзывы: