background image

Chapter 3: Normal Operation & Routine Maintenance

 

 

Product Service 1-800-522-7658 

(6)

 

Place solvent of proper type and quantity into beaker.  A quantity of 
25 to 35 ml of solvent is normally sufficient for an estimate 3 to 4 
hours of refluxing. 

 
(7)

 

Assemble beaker to retainer ring with the wave washer inserted and 
twist to lock securely in place.  After beaker is locked in place, pull 
down elevation control knob and twist ¼ turn to the left.  Raise 
control knob to “catch” position and again twist ¼ turn to the left.  
Slowly allow heater to rise against bottom of flask.  DO NOT let go 
of knob until heater is in position against beaker. 

 
(8)

 

Select heat rate required using panel control switch(es).  Control is 
provided for full range of power with the “HIGH” position providing 
both heaters with approximately 100 watts each.  Control position “5” 
provides approximately 35 watts each. 

 
(9)

 

Turn the main power switch, located on the right side, to the ON 
position.  Switches on the panel each control two heaters. 

 

NOTE: If only one extraction is being made, the adjacent (or 
companion) heater, operating on the same panel switch, should be 
loaded with a beaker of solvent or water to “load” the heater and 
increase its service life. 

 
(10)

 

Pre-heating the heaters for about 8 to 10 minutes prior to starting 
samples will reduce waiting time for start of refluxing. 

 
(11)

 

Water flow rate through the condenser head system can be regulated 
to provide a wide range of reflux rates, when used alone or in 
combination with heater control selections. 

 

NOTE: If water flow rate is reduced too much, solvent will be lost in 
the beakers.  A condenser without flow rate will permit solvent to 
escape through the pressure relief valves. 

 
(12)

 

When extraction is completed, place heater covers on heaters to 
prevent solvent from coming into contact with the hot heater.  
Remove sample tubes, replace with reclaiming tubes and remove 
heater covers.  Heat transfers solvent into the reclaiming tube, leaving 
oil in the beaker. 

 
(13)

 

Drying of the extract is accomplished efficiently with a special swing 
beaker holder (one holder behind each heater assembly) which 
inclines the beaker at a slight angle in the warm air above the heater 
assembly.  When all but the last few milliliters of solvent is 
evaporated, the beaker is placed in a drying oven, weighted and 
compared to the tare weight of the beaker.  Weight difference is the 
amount of extracted material. 

Содержание 3500100

Страница 1: ...Goldfisch Fat Extraction Apparatus Models 3500100...

Страница 2: ...30 a m and 5 30 p m Central Standard Time Part 3518100 Rev L ECO L123 Warranty Labconco Corporation provides a warranty to the original buyer for the repair or replacement of parts and reasonable labo...

Страница 3: ...n 2 CHAPTER 2 INSTALLATION 4 Preparation 4 Location 4 Utility Connections 4 Initial Start Up and Check Out 4 Gasket Assembly 5 CHAPTER 3 NORMAL OPERATION ROUTINE MAINTENANCE 6 Summary of Operation 6 P...

Страница 4: ...have been checked against the following component list and the equipment has been installed and tested As shipped the carton should contain the following Part Number Description 35001 Goldfisch Fat Ex...

Страница 5: ...er to the condenser heads Water exits through the water control valve located on the upper left hand portion of the unit And flows through the manifold assembly depositing cool water in the condenser...

Страница 6: ...Chapter 1 Introduction Product Service 1 800 522 7658 3 Figure 1...

Страница 7: ...ood with proper airflow to handle the extremely flammable solvents that are used during the Goldfisch Extractor s normal operation Utility Connections 1 Once the unit has been placed in its final loca...

Страница 8: ...ing parts package which should be taped to the unit Assemble the gasket wave washer retaining ring and glassware according to assembly drawing instructions as shown in Appendix A Hand tighten the reta...

Страница 9: ...tor Actual test methods sample preparation solvent choices etc are determined by official methods of the A O A C and should be referred to in all cases 1 1 Turn on cold water supply Check visual flow...

Страница 10: ...witch should be loaded with a beaker of solvent or water to load the heater and increase its service life 10 Pre heating the heaters for about 8 to 10 minutes prior to starting samples will reduce wai...

Страница 11: ...h the loss reported as the ether extract Official methods should be referred to for specific test procedure of the indirect method in all cases Maintenance All of the parts contained in the operating...

Страница 12: ...5163 1 Switch infinite control 35046 Pkg of 6 Beaker glass Pyrex tall form 5 high 35053 Pkg of 6 Beaker glass Pyrex 100 cc 3 1 4 high 35054 Ring beaker 35059 Pkg of 6 Extraction thimble 22 x 80 mm 350...

Страница 13: ...Appendix A Replacement Parts Product Service 1 800 522 7658 10...

Страница 14: ...Product Service 1 800 522 7658 11 A Ap pp pe en nd di ix x B B D Di im me en ns si io on ns s 35 7 26 9 9...

Страница 15: ...No 35001 operates on 115 volts 50 60 Hz max at 5 2 amps It is shipped complete with power cord and grounded 3 wire plug Standard features are as follows 1 Main power on off 2 Full range variable heat...

Страница 16: ...Product Service 1 800 522 7658 13 A Ap pp pe en nd di ix x D D W Wi ir ri in ng g D Di ia ag gr ra am m...

Отзывы: