Chapter 1: Introduction 

 

Product Service 1-800-522-7658 

General Description 

The Labconco Goldfisch Fat Extraction Apparatus is designed for single or 
multiple fat extractions according to the methods prescribed in the 
Association of Official Analytical Chemists (AOAC) or other technical 
methods involved with the use of a continuous extractor. 
 
These units operate on a continuous type extraction principle; however, the 
refluxing time normally required by the Soxhlet Extraction Method is usually 
reduced through the design principles incorporated in the Goldfisch 
Extractors. 
 
The following instructions outline the features, installation and maintenance 
of the Goldfisch Extractor.  This manual should be kept handy so that 
questions of use or service can be answered. 
 

Component Identification 

Refer to Figure 1 for the specific location of these components. 
 
1.

 

Manifold Assembly.

  This assembly is located at the top of the apparatus 

and provides an even flow and distribution of water to the condenser 
heads.  Water exits through the water control valve, located on the upper 
left hand portion of the unit.  And flows through the manifold assembly, 
depositing cool water in the condenser head system. 

2.

 

Pressure Release Valve

.  These valves are located at the top of the 

manifold above each of the condensers.  It provides an escape for excess 
gaseous pressure which may build up in the condenser.  The valves 
automatically reset themselves to assure a closed system of extraction. 

3.

 

Condenser Head System

.  These cylindrical shaped condensers located 

above each heater provide an essential part of the extraction process.  The 
solvent vapor rises from the boiling solvent and reaches a point inside the 
condenser head coming into contact with metal cone shaped surface 
which is cooled by water.  The vapors condense and drip through the 
sample held in the ceramic thimble, extracting the fat from the sample. 

4.

 

Heater System

. Each 100 watt adjustable spring loaded heater provides 

an even distribution of heat through the use of coiled heater wires 
embedded in a ceramic material.  The variable heat control switch located 
at the front of the unit provides a variety of settings. 

Содержание 3500100

Страница 1: ...Goldfisch Fat Extraction Apparatus Models 3500100...

Страница 2: ...30 a m and 5 30 p m Central Standard Time Part 3518100 Rev L ECO L123 Warranty Labconco Corporation provides a warranty to the original buyer for the repair or replacement of parts and reasonable labo...

Страница 3: ...n 2 CHAPTER 2 INSTALLATION 4 Preparation 4 Location 4 Utility Connections 4 Initial Start Up and Check Out 4 Gasket Assembly 5 CHAPTER 3 NORMAL OPERATION ROUTINE MAINTENANCE 6 Summary of Operation 6 P...

Страница 4: ...have been checked against the following component list and the equipment has been installed and tested As shipped the carton should contain the following Part Number Description 35001 Goldfisch Fat Ex...

Страница 5: ...er to the condenser heads Water exits through the water control valve located on the upper left hand portion of the unit And flows through the manifold assembly depositing cool water in the condenser...

Страница 6: ...Chapter 1 Introduction Product Service 1 800 522 7658 3 Figure 1...

Страница 7: ...ood with proper airflow to handle the extremely flammable solvents that are used during the Goldfisch Extractor s normal operation Utility Connections 1 Once the unit has been placed in its final loca...

Страница 8: ...ing parts package which should be taped to the unit Assemble the gasket wave washer retaining ring and glassware according to assembly drawing instructions as shown in Appendix A Hand tighten the reta...

Страница 9: ...tor Actual test methods sample preparation solvent choices etc are determined by official methods of the A O A C and should be referred to in all cases 1 1 Turn on cold water supply Check visual flow...

Страница 10: ...witch should be loaded with a beaker of solvent or water to load the heater and increase its service life 10 Pre heating the heaters for about 8 to 10 minutes prior to starting samples will reduce wai...

Страница 11: ...h the loss reported as the ether extract Official methods should be referred to for specific test procedure of the indirect method in all cases Maintenance All of the parts contained in the operating...

Страница 12: ...5163 1 Switch infinite control 35046 Pkg of 6 Beaker glass Pyrex tall form 5 high 35053 Pkg of 6 Beaker glass Pyrex 100 cc 3 1 4 high 35054 Ring beaker 35059 Pkg of 6 Extraction thimble 22 x 80 mm 350...

Страница 13: ...Appendix A Replacement Parts Product Service 1 800 522 7658 10...

Страница 14: ...Product Service 1 800 522 7658 11 A Ap pp pe en nd di ix x B B D Di im me en ns si io on ns s 35 7 26 9 9...

Страница 15: ...No 35001 operates on 115 volts 50 60 Hz max at 5 2 amps It is shipped complete with power cord and grounded 3 wire plug Standard features are as follows 1 Main power on off 2 Full range variable heat...

Страница 16: ...Product Service 1 800 522 7658 13 A Ap pp pe en nd di ix x D D W Wi ir ri in ng g D Di ia ag gr ra am m...

Отзывы: