![Lab companion FHG-150 Скачать руководство пользователя страница 26](http://html1.mh-extra.com/html/lab-companion/fhg-150/fhg-150_operation-manual_3150453026.webp)
23
2)
Turn the
Jog Dial
to select the desired high temperature deviation value.
3)
If selected, press the
Jog Dial
to store it.
4)
To change the low temperature deviation value, press the
MENU Button
again and
check whether three displays (
Temp. Low, Refrig.,
and
Set Temperatue
shown
below inside pink circles) blink along with beep sounds:.
5)
Turn the
Jog Dial
to select the desired low temperature deviation value.
6)
If selected, press the
Jog Dial
to store it.
4.4.5 Temperature Offsetting (Calibration)
The displayed temperature of this equipment can be different from the temperature
measured by your own thermometer which you want to use as standards. In such cases,
you can offset such temperature differences as follows:
1)
Press the
MENU Button
for more than 5 seconds at
Stand-by Mode
and
check
whether the following temperature deviation change display appears along with
beep sounds:
2)
Press the
MENU Button
twice and check whether the following display appears with
three displays (
Temp., BIAS,
and
Set Temperatue
shown below inside pink circles)
blinking: