![L-Card L-502-P-G-D-I Скачать руководство пользователя страница 35](http://html1.mh-extra.com/html/l-card/l-502-p-g-d-i/l-502-p-g-d-i_user-manual_3086592035.webp)
35
4.2: L-502 connectors description.
Signal name
Comm
on
point
Direc-
tion
Description
programmatically.
DAC1
/ +15 V
/ AGND
/ not
connected
(see
section
AGND
Output
For modifications, the L-502-
░
-
░
-D can be configured with a jumper
as the output of the 1st DAC channel (voltage output in the range -5 ...
+ 5 V).
For any modifications, the L-502 can be configured with a jumper as +
15V output of an external device, or as an additional AGND contact, or
as an unconnected pin contact.
(see
section 2.1.2
DAC2
/ -15 V
/ DGND
/ not
connected
(see
section
AGND
Output
For modifications, the L-502-
░
-
░
-D can be configured with a jumper
as the output of the 2nd DAC channel (voltage output in the range -5 ...
+ 5 V).
For any modifications, the L-502 can be configured with a jumper as
an -15V output of an external device or as an additional DGND
contact, or as an unplugged connector pin
(see
section 2.1.2
AGND
—
—
Analog ground
GND32
AGND
Input
•
In the "with common ground" mode: common inverting channel
input 1 ... 32.
•
For all modes must be connected to AGND (in differential mode - to
increase noise immunity). In the "common ground" mode, it is
recommended to connect to the AGND on the signal source side.
•
Operation voltage range ±1 V
(
.
DI_SYN1
DGND
Input
Synchronization input 1, which can also act as an additional input to
the digital input.
Compatible with the output logic level of TTL/CMOS- cells with a
supply voltage of +2.5 V to +5 V. The input has an extended range of
maximum permissible voltages (± 10 V relative to GND).
The minimum rate of rise of the signal drop at the input DI_SYN1 is
not specified, since there is a Schmitt trigger on this input.
There is a software option to turn the 1k pull-up resistor to a high logic
level at this input.
The DI_SYN1 input does not bypass the external TTL source, even
when the power is off.
Notes to
•
The maximum permissible voltages and currents at the contacts of the connectors are indicated in
section