3-5 CONNECTING EXTERNAL CONTROL SIGNALS (Only EDS-400A)
The following operations are enabled by inputting a contact signal or TTL signal (0 to 5 V) to the "CONT.
IN" connector. A given operation is executed when the contact signal turns ON or TTL signal level
becomes L (0 V). The signals thus input must be free of “chatter.”
Note that the EDS-450A is not provided with external control function.
[1] Start/Stop
Inputting a contact signal or TTL signal between C and F of the "CONT. IN" connector, as shown
below, will start recording if it is currently stopped or stop recording if it is currently in progress.
[2] Trigger
Inputting a contact signal or TTL signal between D and F of the "CONT. IN" connector, as shown
below, will issue a trigger when the system is waiting for a trigger in the external trigger mode.
[3] Zero
Inputting a contact signal or TTL signal between E and F of the "CONT. IN" connector, as shown
below, will implement zero suppression.
The CONT
IN connector types and pin assignments
are shown below:
EDS-400A side connector
Type: PRC07-R8F
Manufacturer: Tajimi Electronics
Cable connector
Type: PRC07-P8M
Manufacturer: Tajimi Electronics
Pin No.
Description of signal
A NC
B NC
C Start/stop
input
D Trigger
input
E Zero
input
F GND
G NC
H NC
[Note] NC: Do not connect anything.
Содержание EDS-400A
Страница 3: ...7 SPECIFICATIONS 34 7 1 OVERALL SPECIFICATIONS 34 7 2 HARDWARE SPECIFICATIONS 34 7 3 EXTERNAL VIEW 36 ...
Страница 12: ...9 2 NOMENCLATURE AND MAIN FUNCTIONS 2 1 FRONT EDS 400A EDS 450A 1 2 3 4 5 6 7 10 9 8 1 2 3 4 5 6 7 10 9 8 ...
Страница 15: ...12 2 2 REAR EDS 400A EDS 450A 1 2 5 1 2 3 4 5 ...
Страница 39: ...36 7 3 EXTERNAL VIEW EDS 400A 100 50 110 ...
Страница 40: ...37 EDS 450A ...