11. Battery and Fuse Replacement
#
DANGER
● Never attempt to replace batteries while making measurements.
When replacing the Fuse, use the one with same specification.
#
WARNING
● To avoid electrical shock hazard, disconnect the Test Leads from the
instrument before opening the Battery Compartment Cover. Screw and
fasten the Cover when a replacement work completes.
#
CAUTION
● Do not mix new and old batteries.
Install batteries in the orientation as shown inside the Battery
Compartment, observing correct polarity.
● When disposing the old batteries, please follow your local regulations.
11-1 Battery Replacement
1. Set the Range Switch to the OFF position, and disconnect the Test
Leads from the instrument.
2. Unscrew the Battery-Compartment-fixing screws, and remove the Cover
and replace the batteries with new ones. Replace all 8 batteries.
3. Fix the Cover after replacing batteries and screw up the Cover.
11-2 Fuse Replacement
1. Set the Range Switch to the OFF position, and disconnect the Test
Leads from the instrument.
2. Unscrew the Battery-Compartment-fixing screws, and remove the Cover.
Replace the fuse with new one. (Fig.61)
Fuse Spec : F500mA/ 600V, dia. 6.35 x 32mm.
3. Fix the Cover after replacing a fuse and screw up the Cover.