PT5040 Operation Manual Chapter 6 Introduction to [SYSTEM]
instrument)
For multi-channel instrument, if you select the working type as 2 coil, 3 coil or 4 coil and the
working mode as Normal, the instrument can offer the independent standard waveform with
different voltage and sampling rate for each winding.
Single Std All of the windings use the same standard waveform.
Multi Std All of the windings use the independent standard
waveform with different voltage and sampling rate.
Winding (only for
channel instrument)
It used to set the type of the test windings.
Four-channels instrument
:
Normal Working Mode
:
1 Coil
:
AG
、
BG
、
CG
、
DG
2 Coil
:
AG-BG
、
CG-DG
、
AB-CD
3 Coil
:
AG-BG-CG
、
AB-BC-AC
4 Coil
:
AG-BG-CG-DG
3-Phase 4-Wire
:
AG-BG-CG
3-Phase 3-Wire
:
AB-BC-AC
Balance Working Mode
:
1 Coil
:
Not support
2 Coil
:
AG-BG
、
CG-DG
、
AB-CD
3 Coil
:
AG-BG-CG
、
AB-BC-AC
4 Coil
:
AG-BG-CG-DG
3-Phase 4-Wire
:
AG-BG-CG
3-Phase 3-Wire
:
AB-BC-AC
TWO-channels instrument
:
Normal Working Mode
:
1 Coil
:
AG
、
BG
2 Coil
:
AG-BG
、
AG-BG-AB
Balance Working Mode
:
1 Coil
:
Not support
2 Coil
:
AG-BG
for
channel instrument)
It used to select the current winding, then the impulse voltage and sampling rate
can be set to the correspond to the winding.
5-3