Series 902 Piezo
29
Logarithmic 10
The following graph shows the correlation of logarithmic analog output to
pressure on a 2 to 10 volt scale. This is the output the user will see if the
value is
LOG10
. To receive analog output for
LOG10, the converter must have a 10 volt scale. The logarithmic output is
P (Torr) = 10
(V/2-2)
.
902 Logarithmic Analog Output
(2 - 10 Volt Scale)
0
2
4
6
8
10
0.1
1.0
10.0
100.0
1000.0
Pressure (Torr)
Voltage