2.3 Starting the electronic
Through the Bluetooth-connection the transfer of measured data to the PC is saver,
quicker and easier
– contrary to other devices, which comprehend such data via
parallel interface.
To control the current charge of battery switch on the electronic part with the turning
knob to BAT. This value must always lie over 7.5 Volt (otherwise please charge the
battery). Then turn to the position SENSOR or ECO to read the sensor or calibration
data on the screen. The sensor data should always lie between 1-4, the calibration
approx. 2.5 (before the calibration automatically is set it could temporarily show other
values).