background image

 

 

 
14 

 

    UFM 

3030 

   

Main menu level

Main menu level

Submenu level

Submenu level

Function level

Function level

Data level

Data level

Fct. 4.00.00

PARAM ERR

Fct. 4.01.00

FLOW VELOC

Fct. 4.01.01

FULL SCALE

1.0000E+2

m3/hr

Fct. 1.00.00

OPERATION

Fct. 1.01.00

FLOW

Fct. 1.01.01

FULL SCALE

1.0000E+2

m3/hr

Fct. 1.00.00

STORE YES

Fct. 1.00.00

STORE NO

Fct. 1.00.00

RETURN

Fct. 4.00.00

STORE YES

Fct. 4.00.00

STORE NO

Fct. 4.00.00

RETURN

CODE 2

- -

100.0

m3/hr

Fct. 0.00.01

VIEW ERR

Error

messages

Data acceptance

level

(only appears

after change

of data)

Data acceptance

level

(only appears

after change

of data)

Measuring

mode

Measuring

mode

Error/

Totalizer reset

mode

Error/

Totalizer reset

mode

Parameter

setting

mode

Parameter

setting

mode

Parameter

error

mode

Parameter

error

mode

CODE 1

- - - - - - - - -

Parameter

check

9 x

START

START

 

Key  Measuring mode 

Menu mode 

Data level 

 

Go to the parameter setting mode, 
function 1.00.00 OPERATION. If access 
CODE 1 is activated, CODE 1 must be 
entered first. Using function 3.07.02 
access CODE 1 can be activated or 
deactivated.   

Go to the next, lower 
menu level. 

Go to the next character 
or change line (only 
when 2 lines are 
displayed) 

 

Go to the error/Totalizer reset mode (via 
"CODE 2") 

 

Return to the previous 
(higher) menu level or 
leave the menu mode. 

Accept entered value 

 

Cycle through measured values, see 
Function 3.03.07 CYCL DISP 

Cycle through menu 
options within actual 
menu level. 

Cycle active digit up to 
new values 
 

 

Function  

Text 

Description and settings 

0.00.00 ERROR/TOT 

Main 

menu 

0.00.00 Error/Totalizer 

0.00.01 

VIEW ERR 

View error messages list 

0.00.02 

RST ERR 

Reset error messages 

 

 

NO RESET (keep error messages list) 
RESET (reset error messages list) 

0.00.03 

RST TOTAL 

Reset Totalizer (option available depending on setting in 3.7.8) 
RESET ALL (reset all Totalizer values) 
NO RESET (keep Totalizer values) 

 

 

Содержание UFC 030

Страница 1: ...Heat metering Communications technology Switches counters displays and recorders Engineering systems solutions DIN A4 7 10028 31 00 KROHNE 06 2005 7 30934 32 00 Short form installation and operating...

Страница 2: ...nstall the product securely and stable Install and connect cabling proper to exclude damage or harmful situations If the product does not operate normally refer to the service instructions or refer to...

Страница 3: ...liable in any way for possible consequences of such changes Product liability and warranty Responsibility for suitability and intended use of this ultrasonic flowmeter rests solely with the user Impro...

Страница 4: ...nd your local KROHNE representative A simple operating check of the electronics after unpacking and before permanent installation is advisable to ascertain whether it has suffered damage during shipme...

Страница 5: ...flowmeter by the signal converter housing or the terminal box Check the weight of the flowmeter as indicated on the type plate before handling the unit When handling the flowmeter avoid hard blows jol...

Страница 6: ...entreline and any adjacent wall of at least 0 5 m 1 6 ft Special installation requirements To avoid measuring errors and malfunctioning of the flowmeter due to gas or air inclusions or an empty pipe p...

Страница 7: ...with electric corrosion protection are generally insulated inside and outside so that the fluid has no conductive connection to ground The flowmeter must be insulated from the pipe Note the following...

Страница 8: ...for hazardous areas Power supply connection In case of connection to the mains supply voltage Environmental conditions The UFM3030 is designed to operate safe under the following conditions Suitable f...

Страница 9: ...DC 18 32 V dc AC 24 V AC 10 15 N N1 Neutral power supply PE Protective ground connection FE Functional ground connection Protective conductor clamp terminal Conductors up to 4 mm 11 AWG need to be con...

Страница 10: ...ication Function can be set via menu option 3 4 0 and 3 6 0 Current output I I 22 mA Rload 680 ohm Umax 15Vdc Digital input C low 0 5 VDC high 15 32 VDC Will be switched off when current output activa...

Страница 11: ...the correct converter UFC 030 F is used with the correct flow sensor UFS 3000 Order No see instrument nameplates Meter size DN Function 3 1 5 Primary constant GK Function 3 1 6 Flow direction Functio...

Страница 12: ...rate Velocity of Sound VOS Total Totalizer forward flow Total Totalizer reverse flow Total Totalizer sum and 4 Compass field for error indication 5 Operating keys for programming the signal converter...

Страница 13: ...e required setting or programming task Function block 2 Test contains all available test functions This block can be accessed to check proper functioning off all converter hard and software Function b...

Страница 14: ...easuring mode Menu mode Data level Go to the parameter setting mode function 1 00 00 OPERATION If access CODE 1 is activated CODE 1 must be entered first Using function 3 07 02 access CODE 1 can be ac...

Страница 15: ...IT Pulse value per volume flow unit see Function 3 05 09 1 03 03 PULSE UNIT Pulse value per energy unit see function 3 05 10 2 00 00 TEST Main menu 2 00 00 Test functions 2 01 00 DISPLAY Submenu 2 01...

Страница 16: ...US Gal hr bbls hr bbls day free user configurable unit 3 01 02 ZERO VALUE Zero value see Function 1 01 02 FIXED factory zero setting MEASURED zero calibration possible see Function 3 01 03 3 01 03 ZER...

Страница 17: ...08 through 3 02 11 CORR T P temperature correction via input 1 pressure correction via input 2 see also Function 3 02 08 through 3 02 11 HEAT heat measurement see function 3 02 12 BATCH batch volume s...

Страница 18: ...nction 1 02 02 A Totalizer can be selected here for displaying Available options are TOTAL OFF FORWARD REVERSE BOTH SUM BOTH SUM NO DISPLAY 3 03 04 TOTAL VOL Unit for volume Totalizer X10 m US Gallon...

Страница 19: ...02 11 F R IND forward reverse flow indication VOS velocity of sound range defined in Function 3 01 11 and 3 01 12 DIG OUTPUT digital output see Function 3 05 03 BATCH OUTP batch output indication see...

Страница 20: ...h key only YES entry with key and code 1 factory set on 9 x key see Function 3 07 03 3 07 03 CODE 1 Code 1 Press any 9 keystroke combination and then press the same combination again Each keystroke is...

Страница 21: ...t value for 100 scale see Function 3 04 05 4 03 00 LF CUTOFF Low flow cut off range incorrect If low flow cut off is set to on the value for CUTOFF OFF is compared with the value of CUTOFF ON on Ensur...

Страница 22: ...on 3 04 03 4 09 02 CURR 0 pct Current value for 0 scale see Function 3 04 04 4 10 00 INP OUTP The digital input C and current output I are not allowed to be switched on simultaneously If the Profibus...

Страница 23: ...he liquid used KROHNE regrets that they can not service your flowmeter unless accompanied with such a statement The following specimen statement is available on the KROHNE website as a word file Simpl...

Отзывы: