3.5 

Electrical connection of the signal inputs and outputs 

 
The terminal to connect the electrical signal inputs and outputs consist of 6 connections. 
 
For standard instruments 
 

 

 

For instruments with a communication module 
 

 

P/I/C

 

For wiring of the signal inputs and outputs it is advised to use unshielded twisted pairs. 
 
Internal circuit of the signal inputs and outputs 
of the converter 
 

 

Terminal   Function 

Specification 

 

Common ground  

A1 

Analog input 1, for temperature 
measurement. Configuration via menu 
option 3.2.2. and 3.2.3. 

0(4) to 20 mA 
Ri = 58,2 ohm,  
fuse: 50 mA 

A2 

Analog input 2, for temperature or 
pressure measurement. Configuration via 
menu option 3.2.4. and 3.2.5.  

0(4) to 20 mA 
Ri = 58,2 ohm,  
fuse: 50 mA 

Pulse/frequency output. Function can be 
set via menu option 3.5.0. 

I max: 150 mA 
Umax: 32Vdc, 24Vac 
Max frequency: 2 kHz 

I/C 

Combined Current output (I) and Digital 
input (C). 
Current output (I) Incl. Hart 
communication  
Function can be set via menu option 
3.4.0. and 3.6.0. 

Current output (I): I 

 22 mA,  

Rload 

 680 ohm. Umax = 15Vdc. 

Digital input (C): low = 0-5 VDC, high = 15-
32 VDC. Will be switched off when current 
output activated. 

V+ 

DC power supply from converter for 
active wiring of inputs and outputs 

22 VDC at full load, 24 VDC maximum.  

 100 mA. 

D+ 

Communication con 

For fieldbus communication 

D- 

Communication connection - 

For fieldbus communication 

P/I/C 

Combined current output (I) digital output 
(C) and pulse output (P). See individual  
I/C terminal and P terminal functions 

See individual I/C terminal and P terminal 
specifications. 

The electrical input and output signals can be connected either in active or in passive mode. In 
active mode DC supply voltage is provided from the V+ terminal. In passive mode supply voltage is 
provided from an external source. Please observe instrument polarity: current (I) is always flowing 
towards I, C, P, A1, A2 terminals (current sink). 
Note! Never use the active and passive mode at the same terminal simultaneously. 
If HART communication is used, do not connect the pulse/frequency output P in active mode. 

Содержание UFC 030

Страница 1: ...Heat metering Communications technology Switches counters displays and recorders Engineering systems solutions DIN A4 7 10028 31 00 KROHNE 06 2005 7 30934 32 00 Short form installation and operating...

Страница 2: ...nstall the product securely and stable Install and connect cabling proper to exclude damage or harmful situations If the product does not operate normally refer to the service instructions or refer to...

Страница 3: ...liable in any way for possible consequences of such changes Product liability and warranty Responsibility for suitability and intended use of this ultrasonic flowmeter rests solely with the user Impro...

Страница 4: ...nd your local KROHNE representative A simple operating check of the electronics after unpacking and before permanent installation is advisable to ascertain whether it has suffered damage during shipme...

Страница 5: ...flowmeter by the signal converter housing or the terminal box Check the weight of the flowmeter as indicated on the type plate before handling the unit When handling the flowmeter avoid hard blows jol...

Страница 6: ...entreline and any adjacent wall of at least 0 5 m 1 6 ft Special installation requirements To avoid measuring errors and malfunctioning of the flowmeter due to gas or air inclusions or an empty pipe p...

Страница 7: ...with electric corrosion protection are generally insulated inside and outside so that the fluid has no conductive connection to ground The flowmeter must be insulated from the pipe Note the following...

Страница 8: ...for hazardous areas Power supply connection In case of connection to the mains supply voltage Environmental conditions The UFM3030 is designed to operate safe under the following conditions Suitable f...

Страница 9: ...DC 18 32 V dc AC 24 V AC 10 15 N N1 Neutral power supply PE Protective ground connection FE Functional ground connection Protective conductor clamp terminal Conductors up to 4 mm 11 AWG need to be con...

Страница 10: ...ication Function can be set via menu option 3 4 0 and 3 6 0 Current output I I 22 mA Rload 680 ohm Umax 15Vdc Digital input C low 0 5 VDC high 15 32 VDC Will be switched off when current output activa...

Страница 11: ...the correct converter UFC 030 F is used with the correct flow sensor UFS 3000 Order No see instrument nameplates Meter size DN Function 3 1 5 Primary constant GK Function 3 1 6 Flow direction Functio...

Страница 12: ...rate Velocity of Sound VOS Total Totalizer forward flow Total Totalizer reverse flow Total Totalizer sum and 4 Compass field for error indication 5 Operating keys for programming the signal converter...

Страница 13: ...e required setting or programming task Function block 2 Test contains all available test functions This block can be accessed to check proper functioning off all converter hard and software Function b...

Страница 14: ...easuring mode Menu mode Data level Go to the parameter setting mode function 1 00 00 OPERATION If access CODE 1 is activated CODE 1 must be entered first Using function 3 07 02 access CODE 1 can be ac...

Страница 15: ...IT Pulse value per volume flow unit see Function 3 05 09 1 03 03 PULSE UNIT Pulse value per energy unit see function 3 05 10 2 00 00 TEST Main menu 2 00 00 Test functions 2 01 00 DISPLAY Submenu 2 01...

Страница 16: ...US Gal hr bbls hr bbls day free user configurable unit 3 01 02 ZERO VALUE Zero value see Function 1 01 02 FIXED factory zero setting MEASURED zero calibration possible see Function 3 01 03 3 01 03 ZER...

Страница 17: ...08 through 3 02 11 CORR T P temperature correction via input 1 pressure correction via input 2 see also Function 3 02 08 through 3 02 11 HEAT heat measurement see function 3 02 12 BATCH batch volume s...

Страница 18: ...nction 1 02 02 A Totalizer can be selected here for displaying Available options are TOTAL OFF FORWARD REVERSE BOTH SUM BOTH SUM NO DISPLAY 3 03 04 TOTAL VOL Unit for volume Totalizer X10 m US Gallon...

Страница 19: ...02 11 F R IND forward reverse flow indication VOS velocity of sound range defined in Function 3 01 11 and 3 01 12 DIG OUTPUT digital output see Function 3 05 03 BATCH OUTP batch output indication see...

Страница 20: ...h key only YES entry with key and code 1 factory set on 9 x key see Function 3 07 03 3 07 03 CODE 1 Code 1 Press any 9 keystroke combination and then press the same combination again Each keystroke is...

Страница 21: ...t value for 100 scale see Function 3 04 05 4 03 00 LF CUTOFF Low flow cut off range incorrect If low flow cut off is set to on the value for CUTOFF OFF is compared with the value of CUTOFF ON on Ensur...

Страница 22: ...on 3 04 03 4 09 02 CURR 0 pct Current value for 0 scale see Function 3 04 04 4 10 00 INP OUTP The digital input C and current output I are not allowed to be switched on simultaneously If the Profibus...

Страница 23: ...he liquid used KROHNE regrets that they can not service your flowmeter unless accompanied with such a statement The following specimen statement is available on the KROHNE website as a word file Simpl...

Отзывы: