background image

 

 

 

 

    UFM 

3030 

   21 

3.07.09 

ERR LIMIT 

Error limit in % of measured value for plausibility filter on the sensor 
paths. 

3.07.10 

CNT DECR 

Counter decrement setting for plausibility filter 

3.07.11 

CNT LIMIT 

Counter limit for plausibility filter 
When "0" is set, the plausibility filter will become inactive 
Range: 0 through 1000 
Factory setting: 0 

3.09.00 COMMUNIC Submenu 

3.09.00 Communication 

3.09.01 

PROTOCOL 

Communication protocol  
OFF (no communication) 
HART (HART) 
PROFIB PA (PROFIBUS PA) 

3.09.02 

HART ADDR 

HART address 
Range: 00 through 16 

3.09.03 

PP/FF ADDR 

PROFIBUS PA/FOUNDATION Fieldbus address 
Range: 000 through 126 

 

 

 

4.00.00 

PARAM ERR 

Main menu 4.00.00 Parameter error 

4.01.00 

FLOW VELOC  Volume flow velocity (v) value incorrect. The flow speed is 

calculated from the full scale volume flow and the meter size. 
Ensure condition 0.5 m/s 

 v 

 20 m/s (1,5 to 66 feet/s) is met! 

4.01.01 

FULL SCALE 

Full-scale value for 100% volume flow rate, see Function 3.01.01 

4.01.02 

METER SIZE 

Meter size, see Function 3.01.08 

4.02.00 

CURR OUTP 

Current output range incorrect. Setting for 100% is compared with 
setting for 0%. Ensure condition 100 pct-0 pct 

 4 mA is met! 

4.02.01 

RANGE 

Range of current output, see Function 3.04.03 

4.02.02 

0 pct 

Current value for 0% scale, see Function 3.04.04 

4.02.03 

100 pct 

Current value for 100% scale, see Function 3.04.05 

4.03.00 

LF CUTOFF 

Low-flow cut-off range incorrect: If low flow cut-off is set to on, the 
value for CUTOFF-OFF is compared with the value of CUTOFF-ON 
on. 
Ensure condition CUTOFF-OFF – CUTOFF-ON 

 1% is met! 

4.03.01 

LF CUTOFF 

Low-flow cut-off, see Function 3.01.05 

4.03.02 

CUTOFF ON 

Cutoff “on” value, see Function 3.01.06 

4.03.03 

CUTOFF OFF 

Cutoff “off” value, see Function 3.01.07 

4.04.00 

ENERGY 

Full scale value for heat energy rate (E) incorrect. The fullscale 
value is compared with the maximum value that can be measured 
and should meet the condition: Emax < E fullscale < Emax/1000  
The maximum value that can be measured is at maximum flow and 
200° C temperature difference. 

4.04.01 

HEAT FS 

Full-scale value for 100 % heat energy rate, see Function 3.02.12 

4.05.00 

PULSE/VOS 

Unit of pulse output for velocity of sound function incorrect 
Ensure "PULSE RATE" is selected for "VOS"! 

4.05.01 

PULS FUNCT 

Function of pulse output, see Function 3.05.01 

4.05.02 

PULSE OUTP 

Unit of pulse output, see Function 3.05.07 

4.06.00 

VOS 

Velocity of sound range incorrect: 
Ensure condition MAX VOS 

− 

MIN VOS 

 1 m/s (3.3 feet/sec) is 

met! 

4.06.01 

MIN VOS 

Minimum velocity of sound, see Function 3.01.11 

4.06.02 

MAX VOS 

Maximum velocity of sound, see Function 3.01.12 

4.07.00 

PULSE OUTP 

Pulse output frequency value (f) incorrect. The max frequency is 
calculated from the pulse/unit setting and the max value of the 
measured value. 
Ensure condition 1 pulse/hr 

 f 

 2000 pulse/s is met. 

Содержание UFC 030

Страница 1: ...Heat metering Communications technology Switches counters displays and recorders Engineering systems solutions DIN A4 7 10028 31 00 KROHNE 06 2005 7 30934 32 00 Short form installation and operating...

Страница 2: ...nstall the product securely and stable Install and connect cabling proper to exclude damage or harmful situations If the product does not operate normally refer to the service instructions or refer to...

Страница 3: ...liable in any way for possible consequences of such changes Product liability and warranty Responsibility for suitability and intended use of this ultrasonic flowmeter rests solely with the user Impro...

Страница 4: ...nd your local KROHNE representative A simple operating check of the electronics after unpacking and before permanent installation is advisable to ascertain whether it has suffered damage during shipme...

Страница 5: ...flowmeter by the signal converter housing or the terminal box Check the weight of the flowmeter as indicated on the type plate before handling the unit When handling the flowmeter avoid hard blows jol...

Страница 6: ...entreline and any adjacent wall of at least 0 5 m 1 6 ft Special installation requirements To avoid measuring errors and malfunctioning of the flowmeter due to gas or air inclusions or an empty pipe p...

Страница 7: ...with electric corrosion protection are generally insulated inside and outside so that the fluid has no conductive connection to ground The flowmeter must be insulated from the pipe Note the following...

Страница 8: ...for hazardous areas Power supply connection In case of connection to the mains supply voltage Environmental conditions The UFM3030 is designed to operate safe under the following conditions Suitable f...

Страница 9: ...DC 18 32 V dc AC 24 V AC 10 15 N N1 Neutral power supply PE Protective ground connection FE Functional ground connection Protective conductor clamp terminal Conductors up to 4 mm 11 AWG need to be con...

Страница 10: ...ication Function can be set via menu option 3 4 0 and 3 6 0 Current output I I 22 mA Rload 680 ohm Umax 15Vdc Digital input C low 0 5 VDC high 15 32 VDC Will be switched off when current output activa...

Страница 11: ...the correct converter UFC 030 F is used with the correct flow sensor UFS 3000 Order No see instrument nameplates Meter size DN Function 3 1 5 Primary constant GK Function 3 1 6 Flow direction Functio...

Страница 12: ...rate Velocity of Sound VOS Total Totalizer forward flow Total Totalizer reverse flow Total Totalizer sum and 4 Compass field for error indication 5 Operating keys for programming the signal converter...

Страница 13: ...e required setting or programming task Function block 2 Test contains all available test functions This block can be accessed to check proper functioning off all converter hard and software Function b...

Страница 14: ...easuring mode Menu mode Data level Go to the parameter setting mode function 1 00 00 OPERATION If access CODE 1 is activated CODE 1 must be entered first Using function 3 07 02 access CODE 1 can be ac...

Страница 15: ...IT Pulse value per volume flow unit see Function 3 05 09 1 03 03 PULSE UNIT Pulse value per energy unit see function 3 05 10 2 00 00 TEST Main menu 2 00 00 Test functions 2 01 00 DISPLAY Submenu 2 01...

Страница 16: ...US Gal hr bbls hr bbls day free user configurable unit 3 01 02 ZERO VALUE Zero value see Function 1 01 02 FIXED factory zero setting MEASURED zero calibration possible see Function 3 01 03 3 01 03 ZER...

Страница 17: ...08 through 3 02 11 CORR T P temperature correction via input 1 pressure correction via input 2 see also Function 3 02 08 through 3 02 11 HEAT heat measurement see function 3 02 12 BATCH batch volume s...

Страница 18: ...nction 1 02 02 A Totalizer can be selected here for displaying Available options are TOTAL OFF FORWARD REVERSE BOTH SUM BOTH SUM NO DISPLAY 3 03 04 TOTAL VOL Unit for volume Totalizer X10 m US Gallon...

Страница 19: ...02 11 F R IND forward reverse flow indication VOS velocity of sound range defined in Function 3 01 11 and 3 01 12 DIG OUTPUT digital output see Function 3 05 03 BATCH OUTP batch output indication see...

Страница 20: ...h key only YES entry with key and code 1 factory set on 9 x key see Function 3 07 03 3 07 03 CODE 1 Code 1 Press any 9 keystroke combination and then press the same combination again Each keystroke is...

Страница 21: ...t value for 100 scale see Function 3 04 05 4 03 00 LF CUTOFF Low flow cut off range incorrect If low flow cut off is set to on the value for CUTOFF OFF is compared with the value of CUTOFF ON on Ensur...

Страница 22: ...on 3 04 03 4 09 02 CURR 0 pct Current value for 0 scale see Function 3 04 04 4 10 00 INP OUTP The digital input C and current output I are not allowed to be switched on simultaneously If the Profibus...

Страница 23: ...he liquid used KROHNE regrets that they can not service your flowmeter unless accompanied with such a statement The following specimen statement is available on the KROHNE website as a word file Simpl...

Отзывы: