02/2019 - MA OPTIWAVE-M 7400 C R01 en
45
OPTIWAVE-M 7400 C
OPERATION
4
C
–
Full Setup menu
Menu
No.
Function
Function description
Selection list
Marine Default
C1 Install. Parameters
C1.1
Tank Type
The conditions in which the device is used. If the
surface of the product is flat, select "Storage". If
the surface of the product is disturbed, select
"Process". If the surface of the product is
agitated with vortexes and foam, select
"Agitator". If the device is installed in a stilling
well, select "Stilling Well".
Minimum access level to change the
setting: Expert
Agitator, Stilling Well,
Process, Storage
Storage
C1.2
Tank Height
Tank height is the distance from the flange
face/thread stop of the process connection to
the tank bottom. If you use the device for an
LPR (open-air) application, this value is the
maximum distance that the device must
measure (the measuring range).
Minimum access level to change the
setting: Expert
min-max: 0.0.
100.00
+03
mm /
0.0...2362.2
¨
/
0.0. 196.85 ft
20000 mm
/ 787.402
¨
/ 65.617 ft
1
C1.3
Stilling Well
Height
The height of the stilling well. This menu item is
available if you set "Stilling Well" in menu item
C.1.1 Tank Type.
Minimum access level to change the
setting: Expert
min-max: 0.000.
20.000 m /
0.0...787.40
¨
/
0.0. 65.617 ft
1.0000 m /
39.370
¨
/
3.2808 ft
1
C1.4
Stilling Well
Diameter
The inner diameter of the stilling well. This menu
item is available if you set "Stilling Well" in menu
item C.1.1 Tank Type.
Minimum access level to change the
setting: Expert
min-max: 22.0
.......... 999.0 mm /
866.14
-03
...39.331
¨
/
72.138
-0.3 ....
3.2776 ft
100 mm /
3.937
¨
1
C1.5
Blocking
Distance
The distance from the flange facing or the
thread stop to the top limit of the measuring
range (a zone given by the user where it is not
possible to measure). We recommend a
minimum blocking distance of 100 mm / 4
¨
below the bottom of the antenna. If the
distance is less than the blocking distance, the
device continues to show the blocking distance
on the display screen.
Minimum access level to change the
setting: Expert
min-max: 0.0.
5000.0 mm /
0.0...196.85
¨
/
0.0. 16.404 ft
260 mm
C1.6
Time Constant
Using this function, the device processes several
measurement readings to filter out disturbances.
Increasing the time constant will smoothen the
integrated readings, decreasing will roughen the
readings. s = seconds.
Minimum access level to change the
setting: Expert
min-max:
0. ... 100.0 sec (seconds)
3 s