3
INSTALLATION
24
OPTIWAVE 6300 C
www.krohne.com
04/2011 - 4000305604 - TD OPTIWAVE 6300 R04 en
3.1 Intended use
This radar level transmitter measures distance, level, mass, volume and reflectivity of
granulates and powders.
It can be installed on silos and bunkers.
3.2 Pre-installation requirements
•
Make sure that there is sufficent space on all sides.
•
Protect the signal converter from direct sunlight. If necessary, install the weather protection
accessory.
•
Do not subject the signal converter to heavy vibrations. The devices are tested for vibration
and agree with EN 50178 and IEC 60068-2-6.
3.3 How to prepare the silo before you install the device
3.3.1 Theoretical data for nozzle position
Obey the precautions that follow to make sure that the device is correctly installed.
To avoid measuring errors and device malfunction, obey these precautions.
Follow these recommendations to make sure that the device measures correctly.
Figure 3-1: Recommended nozzle position for solids
1
Position of the process fitting from the silo wall, r/2 (for DN80, DN100 or DN150 horn antennas, and DN80 or DN150
Drop antennas)
2
Radius of the silo, r
3
The minimum measured level for a device without a 2
°
slanted PP flange option
4
The minimum measured level for a device with a 2
°
slanted PP flange option
Содержание OPTIWAVE 6300 C