30
6 Set up with the display and adjustment module
OPTISOUND 3020 C • Two-wire 4 … 20 mA/HART
30507-EN-190318
Sensor-TAG
Sensor
With this menu item, the Basic adjustment is finished and you can
now jump to the main menu with the
[ESC]
key.
In the menu item "
Display
" you can define how the measured value
should be presented on the display.
The following indication values are available:
•
Height
•
Distance
•
Current
•
Scaled
•
Percent
•
Lin. percent
The selection "
scaled
" opens the menu items "
Display unit
" and
"
Scaling
". In "
Display unit
" there are the following options:
•
Height
•
Ground
•
Flow
•
Volume
•
Without unit
Depending on selection, the different units are in turn available.
In the menu item "
Scaling
", the requested numerical value with deci-
mal point is entered for 0 % and 100 % of the measured value.
There is the following relationship between the indication value in the
menu "
Display
" and the adjustment unit in the menu "
Device settings
":
•
Indication value "
Distance
": Presentation of the measured value in
the selected adjustment unit, e.g. m(d).
Displayed value
Scaled
▼
Display unit
Volume
▼
l
▼
Scaling
0 % = 0.0 l
100 % = 100.0 l
A background lighting integrated by default can be adjusted via the
adjustment menu. The function depends on the height of the supply
voltage. See "
Technical data/Voltage supply
".
Display - Indicated value
Display - Backlight