2
INSTALLATION
22
OPTISONIC 1400
www.krohne.com
10/2016 - 4005626301 - QS OPTISONIC 1400 R01 en
•
7
Mount the backend with the transducer onto the valve
•
Check and tighten all
all
all
all the M16 bolts uniformly and in diagonally opposite sequence with a
torque wrench at 14 Nm (12 Nm for the M12 bolts).
•
8
Open the valve, position and align the ultrasonic sensor
Figure 2-26: Installing the device; step 7 and 8