7
TECHNICAL DATA
46
OPTISENS TUR 2000
www.krohne.com
07/2017 - 4001970202 - MA OPTISENS TUR 2000 R01 en
7.2 Technical data
INFORMATION!
•
The following data is provided for general applications. If you require data that is more
relevant to your specific application, please contact us or your local sales office.
•
Additional information (certificates, special tools, software,...) and complete product
documentation can be downloaded free of charge from the website (Downloadcenter).
Measuring system
Parameter
Turbidity
Measuring principle
90
°
scattered light
Design
Measuring principle
Nephelometric
Measuring scale
0...4.000 NTU / FNU
0...40.00 NTU / FNU
0...400.0 NTU / FNU (default)
Sensitivity NTU
70...130% NTU
Zero NTU
±
0.400 NTU all scales
Sensor type
Installation with MAC 100 signal converter or directly to control
system via 4...20 mA
Shaft diameter
60 mm / 2.36"
Shaft length
166 mm / 6.5"
Sensor thread
2" NPT
Hose diameter
6.3...9.5 mm / 1/4...3/8"
Operating conditions
Temperature range
-5...50
°
C / +23...122
°
F
Pressure range
Max. 1 bar at 25
°
C / 14.5 psi at 77
°
F
Measuring range
< 1% of end of measurement scale
Response time
t
90
< 120 seconds (small signal < 3%fs), < 40 seconds
(high signal > 3%fs)
Measuring cycle
2 seconds
Relative humidity
0...95% non condensing
Air pressure cleaning
max. 3 bar / 43.51 psi
Units displayed sensor
NTU (= FNU)
Installation conditions
Ingress conditions
IP68
Weight
Body 420 g / 0.93 lb
10 meter cable 760 g / 1.41 lb
Materials
Sensor body
PVC