OPERATION
5
35
OPTISENS TUR 2000
www.krohne.com
07/2017 - 4001970202 - MA OPTISENS TUR 2000 R01 en
5.3.3 Digital mode
Command Format: ID + M + x
^
For example, if ID = 14, and analog out = ppm type 14M1
^
or 00M1
^
The probe can be configured for digital mode (digital mode=1) or analog 4...20 mA (digital
mode=0).
5.3.4 Analog ouput
Command Format: ID + O + x
^
For example, if ID = 14, and analog out = ppm type 14O1
^
or 00O1
^
The analog output 4...20 mA can be linked to the following scales:
Set the parameter x=1 for the scale 4.000 NTU
x=2 for the scale 40.00 NTU
x=3 for the scale 400.0 NTU
5.3.5 Zero standard setting
Command Format: ID + V + x
^
For example, if ID = 14, and value of the turbidity standard =
0.02 NTU, type 14T0.02
^
or 00T0.02
^
In order to verify if the inserted value has been accepted, digit the command H
Probe response:
ID + M + x
command executed correctly
Probe response:
none
command does not run properly
INFORMATION!
For this Command and for all the following Commands shown in this manual, the answer of the
probe will be a replica of the received Command with <lf> line feed at the beginning and at the
end of the answer.
Probe response:
ID + M + x
command executed correctly
Probe response:
none
command does not run properly
Probe response:
ID + V + x
command executed correctly
Probe response:
none
command does not run properly
INFORMATION!
We suggest to use zero standards close to zero NTU value even if a small adjustment is possible
for high reading values as well.