5
OPERATION
30
OPTISENS TUR 2000
www.krohne.com
07/2017 - 4001970202 - MA OPTISENS TUR 2000 R01 en
Step 7: Sensitivity calibration
Step 8: switching back to measurement
•
Deactivate the function manual hold
manual hold
manual hold
manual hold again.
•
Install the sensor in the measuring location.
Please wait till the measured value is stable.
•
Press
^
^
^
^
to confirm.
Press
or
to change the value and press >>>> to move to the next digit in the displayed value. To select the
standard value (e.g. 100.00 NTU) confirm with
^
^
^
^
.
The submenu sensitivity calibration start calib.
sensitivity calibration start calib.
sensitivity calibration start calib.
sensitivity calibration start calib.is displayed.
•
Press
^
^
^
^
to start the calibration.
The message sensitivity calibration successful
sensitivity calibration successful
sensitivity calibration successful
sensitivity calibration successful is displayed.
•
Press
^
^
^
^
to proceed.
If calibration successful
calibration successful
calibration successful
calibration successful is highlighted, press
^
^
^
^
several times to save the calibration and to return to main
menu level.
INFORMATION!
•
If an error occurs during the calibration procedure, the display shows an error message.
•
If the calibration was not successful please start again and follow the steps mentioned in this
chapter.
•
If the calibration was unsuccessful refer to Troubleshooting on page 40