7.2  Technical data 

INFORMATION!

The following data is provided for general applications. If you require data that is more 
relevant to your specific application, please contact us or your local sales office.

Additional information (certificates, special tools, software,...) and complete product 
documentation can be downloaded free of charge from the website (Downloadcenter).

Measuring system

Parameter

Dissolved Oxygen

Measuring principle

Fluorescence

Design

Measuring principle

Optical, luminophore disk

Measuring scale

0...20 mg/l (ppm)

Sensor type

Installation with MAC 100 signal converter or directly to control 

system via 4...20 mA 

Shaft diameter

60 mm / 2.36"

Shaft length

166 mm / 6.5"

Sensor thread

2" NPT

Hose diameter

6.3...9.5 mm / 1/4...3/8"

Operating conditions

Temperature range

-5...50

°

C / +23...122

°

F

Pressure range

Max. 1 bar at 25

°

C / 14.5 psi at 77

°

F

Measuring range

0...20 mg/l (ppm)

Accuracy

±

 0.1 ppm if < 1.0 %ppm / 

±

 1.0 %sat if < 10.0 %sat

±

 0.2 ppm if > 1.0 %ppm / 

±

 2.0 %sat if > 10.0 %sat

Repeatability

±

0.5% of the scale

Resolution

0.01 ppm

Response time (95%)

large signal (>3% air) 95% < 40 seconds

small signal (<3% air) 95% < 120 seconds

Measuring cycle

8 seconds

Drift

< 1%  year 

Relative humidity

0...95% non condensing

Air pressure cleaning 

max. 3 bar / 43.51 psi

Units displayed sensor

mg/l (ppm) or %

Units displayed converter

mg/l (ppm)

Temperature sensor

Pt100

Installation conditions

Ingress conditions

IP68

Weight

Body 420 g / 0.93 lb

10 meter cable 760 g / 1.41 lb

Содержание OPTISENS ODO2000

Страница 1: ...S ODO 2000 OPTISENS ODO 2000 OPTISENS ODO 2000 OPTISENS ODO 2000 Handbook Handbook Handbook Handbook KROHNE 10 2016 4003689502 MA OPTISENS ODO 2000 R02 en The documentation is only complete when used...

Страница 2: ...y part thereof without the prior written authorisation of KROHNE Messtechnik GmbH Subject to change without notice 2 O www krohne com 10 2016 4003689502 MA OPTISENS ODO 2000 R02 en Copyright 2016 by K...

Страница 3: ...and transport 12 3 4 Configuration of a measuring point 12 3 5 Pre installation requirements 13 3 6 Opening the converter housing 14 3 7 Installation procedure 14 3 8 Sensor use without cleaning func...

Страница 4: ...ion 38 5 3 10 Last calibration date 38 5 3 11 ID value 38 5 3 12 Baud rate 39 5 4 Troubleshooting 39 5 5 Function tables 40 6 Service 41 6 1 Maintenance 41 6 1 1 Cleaning 41 6 1 2 Recalibration 41 6 1...

Страница 5: ...The device fulfils the statutory requirements of the following EC directives EMC Directive 2014 30 EU The manufacturer certifies successful testing of the product by applying the CE marking CAUTION Re...

Страница 6: ...on over the Internet e g when communicating by e mail may involve gaps in security It is not possible to protect such data completely against access by third parties We hereby expressly prohibit the u...

Страница 7: ...tion To prevent any injury to the user or damage to the device it is essential that you read the information in this document and observe applicable national standards safety requirements and accident...

Страница 8: ...e warnings must be observed without fail Even partial disregard of this warning can lead to serious health problems and even death There is also the risk of seriously damaging the device or parts of t...

Страница 9: ...tical disk INFORMATION Inspect the packaging carefully for damages or signs of rough handling Report damage to the carrier and to the local office of the manufacturer INFORMATION Do a check of the pac...

Страница 10: ...itself Figure 2 2 Device description 1 Cleaning nozzle 2 2 NPT 3 Attached cable Figure 2 3 Example for a nameplate 1 Serial number 2 Serial number as barcode 3 Order code 4 Article number 5 Device na...

Страница 11: ...g image INFORMATION Inspect the packaging carefully for damages or signs of rough handling Report damage to the carrier and to the local office of the manufacturer INFORMATION Do a check of the packin...

Страница 12: ...as a 4 20 mA current loop and can be operated with external signal converter or without If operated without the external converter the converter is needed as configuration and calibration tool A compl...

Страница 13: ...scratch the optical disk luminophore with black silicone cover of the sensor Make sure that the optical disk is clean and dust free If necessary clean the optical disk as described on page 41 CAUTION...

Страница 14: ...gnal converter or directly to the control system 2 Configure the measurement range Verify or calibrate sensor if required 3 Mount the sensor into the immersion assembly For further information refer t...

Страница 15: ...9 5 mm 1 4 3 8 in suitable length prepare a connection for the cleaning hose push the hose onto the cleaning connector put the sensor cable and cleaning hose through the adequate extension pipe of the...

Страница 16: ...formation refer to Connecting the sensor to PC on page 21 refer to Connecting the sensor directly to the control system on page 21 and refer to Connecting the sensor cable to the signal converter on p...

Страница 17: ...note the following items to avoid fatal injuries destruction or damage of the device or measuring errors De energise the cables of the power supply before you start any installation works Always keep...

Страница 18: ...ing it down and pulling forwards at the same time 2 and 3 be careful and do not disrupt the retaining band it prevents the cover from getting lost Use a slotted screwdriver with a tip of 3 5 x 0 5 mm...

Страница 19: ...rror current DANGER All work on the electrical connections may only be carried out with the power disconnected Take note of the voltage data on the nameplate INFORMATION Look at the device nameplate t...

Страница 20: ...ion of the sensor cable Connecting the sensor cable to the signal converter Remove the terminal cover Thread the sensor cable through the middle right cable gland 1 Push the wires 7 into terminal J 3...

Страница 21: ...cting the sensor to PC Connect the device to the converter or directly to the control system as stated in above chapters The shield is not connected to the probe but it must be connected to the ground...

Страница 22: ...enus offering the possibility to change presets Measuring mode Main menu Submenu Parameter 3 or 4 pages scrolling with or 2 5 s A quick setup A quick setup A quick setup A quick setup A1 language For...

Страница 23: ...epends on the selected hardware setting in C setup C3 I O To start a simulation process refer to the MAC 100 manual B3 simulation I O B3 6 simulation R3 To start a simulation process refer to the MAC...

Страница 24: ...are setting and the used sensor For further information see function tables C1 1 parameter C1 27 offset C1 28 temperature offset C1 29 product cal C3 I O C3 1 hardware C3 2 current out A C3 3 current...

Страница 25: ...time could be equal to 5 or 10 minutes depending of the difference between the sensor and the ambient temperature The sensitivity calibration is done with the sensor in water vapor saturated air insi...

Страница 26: ...tings In case of measuring in sea water contact the local sales office Perform sensitivity calibration successful Put the sensor into a dark beaker or a cup containing sodium sulfite Na2SO3 oxygen cal...

Страница 27: ...d value After starting up the signal converter the measuring screen appears This is the standard screen which is displayed automatically in the normal operating mode In this mode the calibration can b...

Страница 28: ...ds then release the button You are on the main menu level In the upper line of the display A appears beneath the main menu quick setup quick setup quick setup quick setup is highlighted Press or until...

Страница 29: ...ng operation Restore of the excitation current of the LED of 15 mA Reset the sensitivity and zero point Optical efficiency calibration The sensitivity calibration The probe readout depending on temper...

Страница 30: ...in zero solution Inspect the sensing element If it is damaged proceed to replace it and perform a new sensitivity and zero point calibration Figure 5 2 Zero point calibration 1 Sodium sulfite solutio...

Страница 31: ...The submenu sensitivity calibration sensitivity calibration sensitivity calibration sensitivity calibration is highlighted Press or until the submenu start the calibration start the calibration start...

Страница 32: ...etup B test C setup D service Press to enter the chosen menu You are on the first submenu level In the upper line of the display quick setup and A 1 appear Press or until the submenu product cal produ...

Страница 33: ...t will switch to digital communication Else it will remain in analog operation The calibration and programming via PC gives additional possibilities concerning the signal calibration and is straight f...

Страница 34: ...d conflicts 5 2 6 Sensor mode options As factory setting the sensor is configured in analog operating mode current loop 4 20 mA This is within the programming digital mode 0 Analog mode After 8 second...

Страница 35: ...al mode the operator is able to perform the calibration and to configure the sensor via the RS485 connection The digital mode will be maintained until the next time the sensor is switched on 5 3 Comma...

Страница 36: ...Measurement unit field Measurement unit of the parameter 4 ch aligned on the left 1 space ASCII 32 At the end of the record the probe sends the data of the last calibration than 2 bytes with the BCC...

Страница 37: ...14 and salinity 20000 ppm type 14C200 or 00C200 To check whether the entered value was received type the command A 5 3 6 Atmospheric pressure Command Format ID P x For example if ID 14 and the pressu...

Страница 38: ...10 or 00D13 11 10 The Command is for memorising the last calibration date The operator can use the 8 characters as he likes better no syntaxes limitation 5 3 11 ID value Command Format ID I x For exam...

Страница 39: ...pe otherwise the sun protection will be destroyed Optical disk is damaged Calibrate and if outside of specification replace the optical disk Indicated values are too high Deposits on the optical disk...

Страница 40: ...ppm ID ppm 10 mA for 8 at power on Secondary parameters Salinity chloride 0 600 x 100 ppm 100 ppm step 0 ppm Pressure 500 800 mmHg 760 mmHg Relative humidity 0 100 50 Temperature Compensation range 0...

Страница 41: ...leaned by scrubbing it with a suitable brush Make sure you do not touch or scratch the optical disk 6 1 2 Recalibration It is recommended to check the sensor in regular intervals For this use holding...

Страница 42: ...at no water can get into the sensor which would damage it Replace the optical disk Insert the two rods of the tool into the two small holes of the optical disc Unscrew the old optical disk counter clo...

Страница 43: ...nspection or repair please pay strict attention to the following points Due to statutory regulations on environmental protection and safeguarding the health and safety of the personnel the manufacture...

Страница 44: ...evice when it is returned Date Signature Stamp CAUTION Disposal must be carried out in accordance with legislation applicable in your country Separate collection of WEEE Waste Electrical and Electroni...

Страница 45: ...se an oxygen molecule is getting in contact with the excited fluorophore the energy can also be transferred from the excited fluorophore in a non radiative reaction to the oxygen In that case the inte...

Страница 46: ...ith MAC 100 signal converter or directly to control system via 4 20 mA Shaft diameter 60 mm 2 36 Shaft length 166 mm 6 5 Sensor thread 2 NPT Hose diameter 6 3 9 5 mm 1 4 3 8 Operating conditions Tempe...

Страница 47: ...30 meter 98 43 ft Voltage 9 36 VDC Analog output 4 20 mA loop powered isolated Load 600 Ohm max at 24 VDC Digital output RS485 Approvals CE This device fulfils the statutory requirements of the EC di...

Страница 48: ...L DATA 48 OPTISENS ODO 2000 www krohne com 10 2016 4003689502 MA OPTISENS ODO 2000 R02 en 7 3 Dimensions and weight Dimensions mm Dimensions inch a 60 2 4 b 18 0 71 c 15 0 6 d 100 3 94 e 140 5 51 f 25...

Страница 49: ...NOTES 8 49 OPTISENS ODO 2000 www krohne com 10 2016 4003689502 MA OPTISENS ODO 2000 R02 en...

Страница 50: ...8 NOTES 50 OPTISENS ODO 2000 www krohne com 10 2016 4003689502 MA OPTISENS ODO 2000 R02 en...

Страница 51: ...NOTES 8 51 OPTISENS ODO 2000 www krohne com 10 2016 4003689502 MA OPTISENS ODO 2000 R02 en...

Страница 52: ...ices Head Office KROHNE Messtechnik GmbH Ludwig Krohne Str 5 47058 Duisburg Germany Tel 49 203 301 0 Fax 49 203 301 10389 info krohne com KROHNE 10 2016 4003689502 MA OPTISENS ODO 2000 R02 en Subject...

Отзывы: