5
OPERATION
32
OPTISENS ODO 2000
www.krohne.com
10/2016 - 4003689502 - MA OPTISENS ODO 2000 R02 en
Step 4b: accessing the calibration menu via the main menu quick setup
quick setup
quick setup
quick setup
Step 5: switching back to measurement
•
Deactivate the function manual hold
manual hold
manual hold
manual hold again.
•
Install the sensor in the measuring location.
•
Press >>>> for more than 2.5 seconds, then release the button. You are on the main menu level. In the upper line of the
display "A" appears, beneath the main menu quick setup
quick setup
quick setup
quick setup is highlighted.
MAIN MENU
> A quick setup
> A quick setup
> A quick setup
> A quick setup
B test
C setup
D service
•
Press >>>> to enter the chosen menu.
You are on the first submenu level. In the upper line of the display "quick setup" and "A.1" appear. Press
or
until the submenu product cal.
product cal.
product cal.
product cal. is highlighted.
•
Press >>>> to enter the chosen menu.
The actual oxygen is displayed. Please wait till the value is stable.
There is a count down of 35 seconds for the sensitivity calibration.
Press
^
^
^
^
to confirm the settings.
INFORMATION!
•
If "calibration successful" is highlighted, press
^
^
^
^
several times to save the calibration and to
return to main menu level.
•
If an error occurs during the calibration procedure, the display shows an error message.
•
If the calibration was not successful please start again and follow the steps mentioned in this
chapter.
•
If the calibration was unsuccessful refer to Troubleshooting on page 39