Manual OPTISENS AAM 1050
OPTISENS AAM 1050
8
4.
Installation Procedure
ATTENTION
The sensors are delivered with protective caps to keep them wet. Remove those prior to installation!
Mount the board to a wall at a suitable location. Keep the pipes going to the flow cell as short as possible.
NOTE
For good results the measured water has to be representative for the water that is to be controlled.
Do not place the measurement directly after the injection points, make sure that the chemicals had time to
mix properly with the water and keep all distances short to reduce delay times.
Take the sensors out of their boxes and remove the protective caps. Mount the sensors in the appointed ports and connect the cables
according to their labels.
NOTE
To help you to select the correct sensor: You can recognize Chlorine, Chlorine dioxide and Ozone sensors
by the twin gold rings and Peroxide sensors by their twin platinum rings.
The flow monitor with integrated Pt 100 is much smaller and thinner than the other sensors.
Connect the water inlet and outlet. Make sure that the valve is open and the tap at the bottom of the flow cell is closed.
Open the inlet valve slowly until the water flows through the flow cell with the desired flow rate.
Switch on the power supply for the instrument.
Wait a few minutes until all readings are stable.
For calibration and settings follow all steps described in this manual.