background image

Manual OPTISENS AAM 1050  

 

23 

OPTISENS AAM 1050

 

 

 

 

8.1

 

Configuration of the alarm relay 

 

 

Main menu 

 

 

 

 

 

 

.

 

 

 

 

 

 

 

.

 

 

Basic settings 

 

 

 

 

Basic settings 

 

.

  

 

 

 

 

 

 

.

  

 

Alarm relay 

 

 

 

 

Alarm relay 

  Alarm relay 

 

 

 

 

 

 

  Normally open 

 

 

 

 

 

 

 

  Pulse contact 

 

 

 

 

 

 

 

 

 

 

 

You decide whether the alarm relay should be a Normally Open (NO) or Normally Closed (NC) contact. At 
works it is set to NO, but via the basic settings you can change it to NC. In that case the relay is kept closed 
actively, so in case of power failure it opens, thus issuing an alarm.

 

 

 

 

You can also decide whether the relay should switch permanently during an alarm or whether it should only 
give a pulse. This allows independent acknowledgement of the alarm to shut off horns, for example. It also 
allows registration of subsequent alarms. At works the configuration is permanent contact. 

 
8.2

 

Error messages 

 

Error message  

Cause 

Action 

Slope error 

The slope determined by calibration 
was higher than 200% or lower than 
20 %. 

Please check the sensor connection 
and cable, the flow, and the 
temperature sensor and settings.  
Then repeat the calibration. 
If the error message remains, the 
sensor has to be cleaned, regenerated, 
or replaced. 

Error input 1 

The measuring sensor gives no 
signal. 

Please check the connections, the 
cable and the sensor for signs of 
damage. 

Error input 2  

The temperature sensor gives no 
signal. 

Please check the connections, the 
cable and the sensor for signs of 
damage. This message also appears if 
automatic temperature compensation 
was selected although no temperature 
sensor was used. 

Limit 1 / 2  

The measured value exceeded  
limit 1 (or dropped below limit 2, 
respectively). 

Please check the dosing and readjust 
the control parameters, if necessary. 

Dosage check 1 / 2 

Controller 1 (or 2, resp.) gives out a 
100% output for more than the 
defined period of time. 

Please check the dosing, especially the 
feeding tubes and connections. 

Caution! Carefully check for leaking 
chemicals!

 

 

Ext. controller stop  

The digital input has been short-
circuited (in the „contact“ mode). 

This only indicates the external 
controller stop. If, you have connected 
the flow monitor, this message appears 
due to the „lack of water“ alarm. 

 

Cleaning in progress 
 
 

The automatic sensor cleaning has 
started less than five minutes ago. 

The message will disappear 
automatically five minutes after the 
start of the cleaning process. 
While it´s displayed, calibration is not 
possible. 

 

 

Содержание OPTISENS AAM 1050

Страница 1: ...Installation and Operating Instructions OPTISENS AAM 1050 Measuring and control system for potentiostatic measurements of Chlorine Chlorine dioxide Ozone and Hydrogen peroxide 09 2009...

Страница 2: ...nt 13 6 3 2 Calibration of the temperature measurement 13 6 4 Automatic Sensor Cleaning ASR 14 6 5 Averaging 14 7 Adjustments of the controller 15 7 1 ON OFF controller 16 7 2 P PI controller as impul...

Страница 3: ...s of warranty In addition KROHNE Group s Standard General Conditions of Sale and Delivery found on the back of the invoice and forming the basis of the purchasing contract are applicable General limit...

Страница 4: ...safe usage follow all instructions carefully and pay special attention to all warnings issued in this manual If the device is visibly damaged or has been stored inappropriately or if there are any do...

Страница 5: ...N OFF controller with hysteresis P controller as Pulse Pause Impulse Frequency or steady controller PI controller as Pulse Pause Impulse Frequency or steady controller Hysteresis 0 00 4 00 mg l or 0 0...

Страница 6: ...ia 4 predrilled holes Water connection Connect the water inlet on the left side and the outlet on the right side of the flow cell The outlet can be an open outlet pressureless or the water can be re d...

Страница 7: ...tor C Resistance R 60 mA 10 nF 260 V 390 Ohm 2 Watt 70 mA 47 nF 260 V 22 Ohm 2 Watt 150 mA 100 nF 260 V 47 Ohm 2 Watt 1 0 A 220 nF 260 V 47 Ohm 2 Watt 3 1 Connection diagram Connection Terminals Notes...

Страница 8: ...t of their boxes and remove the protective caps Mount the sensors in the appointed ports and connect the cables according to their labels NOTE To help you to select the correct sensor You can recogniz...

Страница 9: ...ring With key Enter changes are stored For your convenience arrows in the display indicate the directions you can take from your position in the menu 5 1 How to adjust parameters Temp comp Manual 1 Wh...

Страница 10: ...s Temp comp Temperature compensation only for pH measurements Enter password Password function Set points Controller settings set points P ranges I functions Limit values Alarm function Basic settings...

Страница 11: ...asic settings Cal Pt Calibration of the temperature measurement Averaging Lack of water indication or flow measurement Temp coeff Activation of averaging function Cleaning Coefficient for linear temp...

Страница 12: ...rature sensor is done at the factory and cannot be changed after shipping 6 2 Calibration Calibration of the measurement of Cl2 ClO2 O3 and H2O2 Main menu Calibration DPD Calibration DPD Calibration 0...

Страница 13: ...e can be regarded as constant you can enter the temperature manually instead of measuring it continuously The device will then compensate the temperature error of this temperature 6 3 1 Temperature co...

Страница 14: ...sible and after cleaning the sensor needs some time for polarisation Therefore the display and the analog output will show the last measured value for five minutes and the message cleaning in progress...

Страница 15: ...red value exceeds the set point and OFF if it drops back below it or vice versa Dosage is always carried out with 100 ON or 0 OFF The parameter for an ON OFF controller is the hysteresis P controller...

Страница 16: ...0 01 mg l For an ON OFF controller you have to set the following parameters 1 Set points S1 and S2 Set point S1 refers to relay 1 set point S2 refers to relay 2 2 P range and integral action time for...

Страница 17: ...irection S1 Basic settings Raise Basic settings Direction S2 Contr settings lower For an impulse frequency controller you have to set the following parameters 1 Set points S1 and S2 S1 refers to relay...

Страница 18: ...a pulse pause controller you have to set the following parameters 1 Set points S1 and S2 S1 refers to relay 1 S2 refers to relay 2 2 P range and integral action time Adjust a P range 0 For a P control...

Страница 19: ...xternal controller stop You can activate or deactivate the controller with an external switch by using the digital input This feature can also be used as low water indication Just connect a level or f...

Страница 20: ...1 402 mS 25 0 C S1 S2 Man 6 Switch ON relay 2 with key The square to the left of S2 gets dark 1 402 mS 25 0 C S1 S2 Man 7 Switch OFF relay 2 again with key The square gets light 1 402 mS 25 0 C S1 S2...

Страница 21: ...inguished NOTE If you work in manual mode Man and you want to prevent that an alarm is triggered during maintenance please check the HOLD Function see chapter 8 Operation and maintenance 7 8 1 Turn on...

Страница 22: ...cause of alarm is eliminated Sensor check during calibration If a sensor gives unsatisfying results during calibration an alarm is issued The alarm is held until a new calibration shows satisfying sen...

Страница 23: ...ensor has to be cleaned regenerated or replaced Error input 1 The measuring sensor gives no signal Please check the connections the cable and the sensor for signs of damage Error input 2 The temperatu...

Страница 24: ...4 20 mA the resolution is lower but defective cable connections are immediately evident With the parameters Begin and End you define which part of the measuring range you want to read out Alternativel...

Страница 25: ...by a Bayonet lock Turn the lock to the left until the fuse pops up Exchange it and fix the new fuse by turning the lock to the right Put the front back on and fix it tightly WARNING Disconnect the pow...

Страница 26: ...a sensor make sure that the replacement sensor is an original KROHNE product and that it corresponds to the equipment used Please mind that you have to calibrate whenever you change a sensor or an in...

Страница 27: ...ation of the device hardware and software 10 9 2 Analog inputs Here you can see the raw data the device obtains from the sensors They are not influenced by compensations or calibration and offer valua...

Страница 28: ...vanically isolated max loading 500 Ohm Interface RS485 Baud rate 9600 data format 8Bit 1 Start and 1 Stop bit Controller ON OFF controller with hysteresis P or PI controller as Pulse Pause or Impulse...

Страница 29: ...nfirming that the device is safe to handle If the device has been operated with toxic caustic flammable or water endangering products you are kindly requested to check and ensure if necessary by rinsi...

Страница 30: ...tputs Output 1 0 20mA 4 20mA for conductivity Controller S1 Controller S2 Start End Controller Controller S1 Controller S2 Acting direction Acting direction raise lower raise lower Set point Set point...

Отзывы: