.52+1(
BM 70 A Kurzanleitung
Seite: 18
4. Entwurf 03/97
Operating temperature at flange
Horn antenna: -30°C to +130°C (T
amb
< 50°C)
(-22°F to +266°F (T
amb
< 122°F)
or -30°C to +120°C (T
amb
< 55°C)
or(-22°F to +248°F (T
amb
< 131°F)
High temperature version:
FFKM: -30 °C to +250°C (-22°F to +482°F)
Kalrez 2035: max. 210°C ( 410°F )
Viton or FEP-coated: max. 200°C (392°F)
Wave-Stick: -20°C to +150°C (-4°F to +302°F), pressure-
dependent
Safety limit
(FFKM): 280°C (536°F)
Product temperature
Unrestricted, provided ambient temperature and flange
temperature are within the specified limits
Ambient temperature T
amb
-20°C to +55°C (-4°F to + 131°F)
(Signal converter)
(with Wave-Stick: max. 50°C (122°F))
Power supply
24 V DCAC
18 - 31.2 V DC or 18 - 26.4 V AC (45 - 66 Hz)
115/230 V AC
85 - 127 V AC or 170 - 254 V AC (45 - 66 Hz)
Power consumption (typical)
approx. 10 W; AC: approx. 12 VA
Microwaves
Measuring principle
FMCW Radar
Frequency range
X-Band 8.5 - 9.9 GHz
Antenna radiation angle
Type 3: ± 8°
Type 4: ± 6°
Wave-Stick: ± 9°
Ex-e current output HART
®
(active)
Current
4 - 20 mA; without or with error message 2 mA or 22 mA
Accuracy and linearity
0.05 %
Load impedance
< 500 ohms
Digital communication
HART
®
(point-to-point)
Switching output (passive)
(only in conjunction with current output HART
®
)
Contact type
1 N/O (electronic)
Contact rating
max. 30 V DC / 30 V AC; 100 mA
Digital input
(only in conjunction with current output HART
®
)
External voltage
5 to 28 V DC
Internal resistance
> 1 kohm
Ex-i current output HART
®
(passive)
Current
4 - 20 mA; without or with error message 22 mA
Accuracy and linearity
0.05 %
Supply voltage
min. 8 V, max. 30 V
Digital communication
HART
®
(point-to-point or multidrop)
RS 485 interface
Transmission rate
1200, 2400, 4800, 9600, 19200 or 38400 bauds
Protocols
Krohne protocol; ModBus RTU, HART
Additional current output
4 - 20 mA active (load impedance < 250 ohms)
Ambient conditions
Environment class
Locations exposed direct to open-air climate,
D1 Severity in conformity with EN 60654-1
Protection category (signal converter) IP66 / IP67 (equivalent to NEMA 4 and 4X)
Electrical connection
Cable entries:
3 x M25×1.5
Terminals:
Cable cross-section 0.5 - 2.5 mm² (solid conductor: max. 4
mm²)
U-clamp terminals
(for PA and PE/FE) cable cross-section max. 4 mm²