Operation & Installation Manual V12-V6 2010-05-31.doc
31/05/2010
46/60
To restart the logging process:
Open the “Monitoring” menu
Click “Logging”
The check mark will reappear and the status field will display “Log.On” and turn to green
again.
Logged data will be stored in a file for each day (provided it is the same meter).
The software will automatically assign a name to file where the logged data will be stored,
the date will be part of the file name (see also Figure 6-13)
When the logging process is restarted - after it has been interrupted – the new data will be
appended to the same file as used before (assuming this will happen on the same day).
When a new day breaks a new log file will be created even while a logging process is
running.
Logged data will be stored in a file with “.CSV” format.
The file will be stored in the folder: C:\KrohneData\Logging\...........
6.4 Adjusting Meter factor (For Authorised personnel only)
A typical adjustment of the parameter settings is the adjustment of the meter factor, This
meter factor is determined during calibration and the meter factor can be installed before
sealing the Ultrasonic flow meter. This is however dependant on the requirements of the
customer.
Adjustment of the meter factor for CT applications can only be done under supervision of a
calibrating officer.
Open the object tree details (view)
Open the tree by clicking on “+”
Figure 6-24: Opening object tree
Go to (the numbering within the tree may differ from the example below):
FlowCalculationModule > Configuration > Flow Calcs > MeterConstant_Fwd