DCM8500PV
Smart Hold
The meter will beep continuously and the display will flash if the measured signal is
larger than the display reading by 50 counts.
(However, it can not detect across the AC and DC Voltage / Current).
AC
AC
HOLD
ZERO
A
A
HOLD
HOLD
DCA ZERO
Remove the jaw from the conductor before performing DCA ZERO.
Press HOLD button for over 2sec. to compensate the residual magnetism.
APO
DC
A
Zero
APO
DC
A
Zero
APO
HOLD
ZERO
> 2 sec
Press
HOLD
ZERO
> 2 sec
Press
DC
A
High Frequency Rejection (HFR)
The High Frequency Rejection mode equip a low pass filter in the AC measurements.
The cut-off frequency (-3dB point) of low pass filter is 800Hz.
HFR
HFR
AC
AC
HFR
Press
HFR
V
V
!
Warning
The hazardous voltage may be present even if the LCD reading is very low in HFR
mode. Verify the voltage again without HFR mode
08