(12) Frequency
Minimum Sensitivity :
> 5V (for ACV 1Hz ~ 10kHz)
> 8A (for ACA 1Hz ~ 1kHz)
Minimum Frequency :
1Hz
Overload Protection :
AC/DC 1000V and 600A
(13) HFR (High Frequency Rejection)
Available for ACV, ACA and, Flexible Current Probe.
Add ± 4% to specified accuracy of each function and each range for 45Hz to 200Hz.
Accuracy is unspecified for > 200Hz.
Cut-off Frequency (-3dB) : 800Hz
(14) Inrush Current
Available for ACA and Flexible Current Probe.
Trigger level :
≧
50d.
Add ± 3% to specified accuracy of each function and each range.
Range
OL Reading
Resolution
Accuracy
100.00Hz
100.00Hz
0.01Hz
1000.0Hz
1000.0Hz
0.1Hz
10.000kHz
10.000kHz
0.001kHz
±(0.3% + 3D)
(15) Resistance
To get more accuracy result, short the test probes to obtain the offset. The accuracy
specification is specified for the result that the offset is subtracted.
Overload Protection :
AC/DC 1000V
(16) Continuity
Built-in buzzer sounds when measured resistance is less than 20Ω and sounds off
when measured resistance is more than 200Ω, Between 20Ω to 200Ω the buzzer
maybe sound or off either.
Continuity Indicator :
2.7K Tone Buzzer
Response Time of Buzzer :
< 100msec.
Overload Protection :
AC/DC 1000V
Range
OL Reading
Resolution
Accuracy
600.0Ω
660.0Ω
0.1Ω
±(0.9% + 5D)
6.000kΩ
6.600kΩ
0.001kΩ
600.0kΩ
660.0kΩ
0.1kΩ
60.00kΩ
66.00kΩ
0.00kΩ
±(0.9% + 2D)
Open Circuit Voltage :
Approx. 1.8V
Overload Protection :
AC/DC 1000V.
(17) Diode
Range
OL Reading
Resolution
Accuracy
1.500V
1.550V
0.001V
±(0.9% + 2D)
16