Symbols as marked on the Meter and Instruction manual

!

Safety Information

Understand and follow operating instructions carefully. Use the meter only as.

WARNING

!

If the equipment is used in a manner not specified by the manufacturer, the 
protection provided by the equipment may be impaired.
Always use proper terminals, switch position, and range for measurements.
To reduce the risk of fire or electric shock, do not use this product around 
explosive gas or in damp locations.
Verify the Meter operation by measuring a known voltage. If in doubt, have the 
Meter serviced.
Do not apply more than the rated voltage, as marked on Meter, between 
terminals or between any terminal and earth ground.
To avoid false readings that can lead to electric shock and injury, replace the 
battery as soon as low battery indicator blinks.
Avoid working alone so assistance can be rendered.
Do not use the Tester if the Tester is not operating properly or if it is wet.
Individual protective device must be used if hazardous live parts in the installation 
where the measurement is to be carried out could be accessible.
Disconnect the test leads from the test points before changing the position of the 
function rotary switch.
Never connect a source of voltage when the function rotary switch is not in 
voltage position.
When using test leads or probes, keep your fingers behind the finger guards.
Use caution with voltages above 30 Vac rms, 42 Vac peak, or 60 Vdc. These 
voltages pose a shock hazard.
Remove test lead from Meter before opening the battery door or Meter case.
DO NOT USE the test leads when the internal white insulation layer is exposed.
DO NOT USE the test leads above maximum ratings of CAT.
environment, voltage and current, that are indicated on the
probe and the probe tip guard cap.
DO NOT USE the test leads without the probe tip guard cap in CAT III and CAT IV 
environments.
Probe assemblies to be used for MAINS measurements shall be RATED as 
appropriate for MEASUREMENT CATEGORY III or IV according to IEC 
61010-031 and shall have a voltage RATING of at least the voltage of the circuit 
to be measured.
Disconnect circuit power and discharge all high-voltage capacitors before testing 
resistance, continuity, diodes, or capacitance.
De-energize the installation under test or wear suitable protective clothing during 
fitting and removal of the Flexible Current Probe.
Do not apply around or remove from UNINSULATED HAZARDOUS LIVE 
conductors, which may render electric
shock, electric burn, or arc flash.

!

Risk of electric shock

Equipment protected by 
double or reinforced 
insulation

DC measurement

Low battery

Earth ground

AC measurement

Both direct and alternating current

Wireless transmission

Conforms to EU directives

Application around and 
removal from hazardous 
live conductors is permitted

See instruction manual

~
~

Do not discard this product or 
throw away.

01

Содержание DCM8500PV

Страница 1: ......

Страница 2: ...probes keep your fingers behind the finger guards Use caution with voltages above 30 Vac rms 42 Vac peak or 60 Vdc These voltages pose a shock hazard Remove test lead from Meter before opening the battery door or Meter case DO NOT USE the test leads when the internal white insulation layer is exposed DO NOT USE the test leads above maximum ratings of CAT environment voltage and current that are in...

Страница 3: ...gital display 7 Input Terminal for Multi function 8 Common Ground reference Input Terminal Cleaning Periodically wipe the case with a dry cloth and detergent Do not use abrasives or solvents Introduction The Meter Description Unsafe Voltage To alert you to the presence of a potentially hazardous voltage when the Tester detects a voltage 30 V or a voltage overload OL in V mV LoZ The symbol is displ...

Страница 4: ...suring function Dial the switch and press the Function button to select AC DC mode V APO AC V APO APO DC DC mV COM Measuring Voltage Measuring PV Voltage V APO AC V APO DC PV String View side COM CAUTION This function is only available with the dedicated PV test probe Always select correct DC AC mode to perform high voltage measurement This meter will flash symbol and the correct mode symbol AC DC...

Страница 5: ...for hand guard Do not hold the meter across the Barrier COM Dial the switch and press the Function button to select the measuring function Note Torch will be on when jaw is opened Dial the switch and press the Function button to select the measuring function Measuring Resistance Continuity Capacitance Diode APO APO k APO Volt seek APO V APO F COM 04 ...

Страница 6: ...ch and press the Function button to select C F mode Measuring Temperature C F APO APO C F Funtion Button COM Dial the switch and press the Function button to select AC DC mode Measuring μA DC A Flame sensor Control Module COM 05 ...

Страница 7: ...lamp meter at 3000A 3V output ratio Measuring Current with Flex Clamp Meter Press the Function button to change the function on the same switch position Using the Function Measuring Frequency Function Position Switch Funtion Button AC AC V Hz Hz 06 ...

Страница 8: ... When the input goes below the recorded min value or above the recorded max value the meter records the new value Press Hold button to pause the recording Press MIN MAX button for over 2 sec to enter exit Volt Seek mode Press MIN MAX button to switch high low sensitivity Warning The Volt Seek LED indicates the electric field If the Volt Seek LED is not on voltage could still be present MIN MAX MIN...

Страница 9: ...rming DCA ZERO Press HOLD button for over 2sec to compensate the residual magnetism APO DC A Zero APO DC A Zero APO HOLD ZERO 2 sec Press HOLD ZERO 2 sec Press DC A High Frequency Rejection HFR The High Frequency Rejection mode equip a low pass filter in the AC measurements The cut off frequency 3dB point of low pass filter is 800Hz HFR HFR AC AC HFR Press HFR V V Warning The hazardous voltage may...

Страница 10: ...HFR INRUSH button before triggering the inrush current measurement 1 2 A A A HFR 2 sec Press Watting Tigger After Tigger 1 2 Inrush 2 Sec HFR 2 sec Press Inrush 2 Sec AC AC AC Auto Power Off Wake up the meter by dialing the switch or pressing any button After Specified Time Without Operation 09 ...

Страница 11: ... 5 minutes 10 minutes 20 minutes and disabled OFF To turn on the meter after keeping HOLD button down Funtion Button 1 2 3 Funtion Button APO HOLD 1 2 APO ZeroMAXMIN HOLDAC LOG HFR LOZ DC S C F FA mV Hz k Function of LOG Button Pressing button while powering up to select the mode Data Logger mode Manual Saving mode and Clear memory Press 1 2 3 Press LOG LOG LOG Press Press 10 ...

Страница 12: ... Press Press LOG LOG S S V AC LOG Press Time interval setting mode Start to record auto matically after 2sec without operaction LOG icon flashes while logging LOG 2 Sec 2Sec CAUTION All stored data will be cleared next startup Download the stored data by App first if needed Manual Saving Mode CAUTION All stored data are saved until switching to data logger mode or executing the clear function LOG ...

Страница 13: ... up to 10m Download KPS Link App via the following QR Code Turn on Bluetooth function of the meter and open KPS Link to connect the DMM The Bluetooth icon of the meter will freeze on LCD after the connection establishes successfully AC Press AC V V Default Temperature Units Setting Turn on the meter after keeping MIN MAX button down C F OR 1 2 COM MIN MAX MIN MAX 12 ...

Страница 14: ...age will flash Power Requirement AA Size Battery x 2 R6 LR6 15D 15A Battery Life 200 hours ALKALINE Battery without Backlight Operating Temperature 10 10 C 10 C 30 C 80 RH 30 C 40 C 75 RH 40 C 50 C 45 RH Storage Temperature 20 C to 60 C 0 to 80 R H batteries not fitted Altitude 6561 7 ft 2000m Safety EN 61010 1 EN 61010 2 032 EN 61010 2 033 for CAT III 1000V CAT IV 600V EN 61326 1 Drop Protection ...

Страница 15: ...curacy is given as of reading counts of least significant digit at 23 C 5 C with relative humidity Less than 80 R H and is specified for 1 year after calibration 3 DC mV SInput Impedance 10MΩ Overload Protection AC DC 1000V Range OL Reading Resolution Accuracy 600 0mV 660 0mV 0 1mV 0 7 5D 4 DC Voltage SInput Impedance 10MΩ Overload Protection AC DC 1000V Range OL Reading Resolution Accuracy 600 0V...

Страница 16: ...y The measured value 1000A add 0 5 to the accuracy Overload Protection AC DC 600A Range OL Reading Resolution Accuracy 60 00A 66 00A 0 01A 600 0A 660 0A 0 1A 2 0 5D 10 AC Current Add 10 dgt to the accuracy when 5 0A Add 0 5 to the accuracy when 1000A Add 1 to the accuracy when 100Hz Frequency Response Sine Wave 45 400Hz Overload Protection AC DC 600A Range OL Reading Resolution Accuracy 60 00A 66 ...

Страница 17: ...Hz 10 000kHz 0 001kHz 0 3 3D 15 Resistance To get more accuracy result short the test probes to obtain the offset The accuracy specification is specified for the result that the offset is subtracted Overload Protection AC DC 1000V 16 Continuity Built in buzzer sounds when measured resistance is less than 20Ω and sounds off when measured resistance is more than 200Ω Between 20Ω to 200Ω the buzzer m...

Страница 18: ...verload Protection AC DC 1000V 19 VoltSeek Voltage Range of High Sensitivity 80V 1000V At the top edge of the jaw Voltage Range of Low Sensitivity 160V 1000V At the top edge of the jaw The accuracy does not include the accuracy of the thermocouple probe Accuracy specification assumes surrounding temperature stable to 1 C For surrounding temperature changes of 2 C rated accuracy applies after 2 hou...

Страница 19: ...PV Test Leads Instruction Probe tip guard cap For CAT III or CAT IV environments use the test leads with the probe tip guard cap fixed firmly Without the probe tip guard cap the test leads can be used in CAT II environment ONLY For 1500V AC 2000V DC measurement This test lead can only be used in the environment that is not connected to MAINS directly Finger Guard 18 ...

Страница 20: ...attempt to repair this test lead set It contains no user serviceable parts Repair or servicing should only be performed by qualified personnel Cleaning Clean the test lead with a water and mild detergent DO NOT use abrasives or solvents and DO NOT IMMERSE in liquid Specification Input Impedance 10MΩ Overvoltage Category CAT 0 1500V AC 2000V DC CAT II 1000V AC 1500V DC CAT III 1000V CAT IV 600V Pol...