CardioCoach PRO User Manual | Page 97
Ambient Temperature Specifications
Absolute Maximum Ratings
Maximum Readings
75 °C
Operating Characteristics
Accuracy
±2 °C
Resolution
1 °C
Minimum Range
10 °C
Maximum Range
40 °C
Frequency Response
t
10-90
< 10 sec
Sample Rate
> 1 Hz
VO
2
Specifications
Operating Characteristics
Accuracy
greater of ±10% Reading or ±10 ml/min O
2
Resolution
1 ml/min O
2
Minimum Range
70 ml/min O
2
Maximum Range
4000 ml/min O
2
FeO
2
Specifications
Operating Characteristics
Accuracy
greater of ±2% of Reading or ±0.2% O
2
Resolution
0.01% O
2
Minimum Range
10% O
2
Maximum Range
30% O
2
Tidal Volume Specifications
Operating Characteristics
Accuracy
greater of ±5% of Reading or ±10 ml
Resolution
1 ml
Minimum Range
200 ml
Maximum Range
4000 ml