Starting the software

 

Start the software and set it for control of the CM-3600A.

Turning the power ON 

Turn the power ON. (P. 15)

Starting the PC 

Turn on the PC to be used to control 

(starting Windows)

 

the CM-3600A.

Attaching a target mask

Attach the target mask 
to be used. (P. 16)

Attach the LAV (Ø25.4 mm) 
target mask and set the white 
calibration plate in position. (P. 16)

Performing zero calibration

Position the zero calibration 
box and perform zero 
calibration. (P. 17)

Position the Transmittance 
Zero Calibration Plate and 
perform zero calibration. (P. 19)

Performing white (100%) 
calibration

Position the white calibration 
plate and perform white 
calibration. (P. 18)

Perform 100% transmittance 

with nothing or a cell 

containing distilled water in the 

transmittance chamber. (P. 20)

Positioning a specimen 

Position the specimen on / in the CM-3600A. (P. 21)

Performing measurement 

Perform measurements.

Turning the power OFF 

When measurements are complete, turn the power OFF. (P. 15)

 

Exit out of the software and turn off the PC.

Connecting the PC

            

Connect the CM-3600A to the PC with the USB cable.

  

 

(P. 13)

Connecting the AC adapter

 

Connect the CM-3600A to the AC outlet 

with the AC adapter. (P. 14)

<Reflectance measurement>

<Transmittance measurement>

12

Measurement Procedure

 

This manual explains how to prepare the CM-3600A and how to set a specimen.

 

The CM-3600A is controlled by a PC to perform measurements.

 

For a description of measuring method using SpectraMagic™ NX 
(optional), refer to the SpectraMagic™ NX instruction manual.

 

Flow of Preparation and Measurement

Содержание CM-3600A

Страница 1: ...SPECTROPHOTOMETER CM 3600A INSTRUCTION MANUAL Before using this instrument please read this manual...

Страница 2: ...le the instrument Denotes an instruction Be sure to disconnect the AC adapter from the AC outlet This symbol indicates A C This symbol indicates D C This symbol indicates class II protection against e...

Страница 3: ...t is damaged or the AC adapter is damaged or if smoke or odd smells occur Doing so may cause a fire In such situations turn the power OFF immediately disconnect the AC adapter plug from the AC outlet...

Страница 4: ...y can be used to protect the instrument from impact and vibration during shipment to Konica Minolta for maintenance Should they be lost or damaged contact the nearest Konica Minolta authorized service...

Страница 5: ...he instrument Measurement Make sure no dirt or dust get into the specimen measuring port When using the instrument for long periods of time the measurement value may change depending on changes in the...

Страница 6: ...is not in use be sure to attach a target mask to the measuring port to prevent entry of foreign matter The CM 3600A may cause interference if used near a television radio etc Since the CM 3600A uses...

Страница 7: ...not in use so that the White Calibration Plate is not exposed to ambient light during storage The Target Masks may discolor if they are left exposed to light When they are not in use keep them in a sa...

Страница 8: ...forming 100 calibration 20 Setting up instrument for 100 calibration to air 20 Setting up instrument for 100 calibration to water 20 Setting a Specimen 21 Removing the Sample Holder 21 Reflectance mea...

Страница 9: ...7 Using the CM 3600A...

Страница 10: ...illumination aperture area at specimen surface areas for each target mask are as follows CM A105 LAV 25 4 mm 30 mm CM A106 MAV 8 mm 11 mm CM A107 SAV 4 mm 7 mm Zero Calibration Box CM A104 Used to pe...

Страница 11: ...c Cell CM A130 optical path length 2mm CM A131 optical path length 10mm CM A132 optical path length 20mm This disposable plastic container is used to hold liquid specimens Cell CM A97 optical path len...

Страница 12: ...mm illumination Transmittance Specimen Holder CM A96 CM A100 Zero Calibration Box CM A104 CM 3600A Target Mask CM A106 MAV 8 mm measurement 11 mm illumination Target Mask CM A105 LAV 25 4 mm measurem...

Страница 13: ...the CM 3600A Sample holder Used to hold the specimen white calibration plate or zero calibration box Power switch Used to turn power ON and OFF Transmittance chamber cover Covers the transmittance cha...

Страница 14: ...ransmittance with nothing or a cell containing distilled water in the transmittance chamber P 20 Positioning a specimen Position the specimen on in the CM 3600A P 21 Performing measurement Perform mea...

Страница 15: ...onnect the USB cable connector check the shape of the receptacle connection terminal and insert the connector fully until it is secured Connecting a Personal Computer Operating Procedure In general a...

Страница 16: ...e electric shock Insert the power plug fully and securely Incomplete inserting may cause fire or electric shock Do not disassemble or modify the AC adapter Doing so may cause a fire or electric shock...

Страница 17: ...fire In such situations turn the power OFF immediately disconnect the AC adapter from the AC outlet and contact the nearest Konica Minolta authorized service facility Turning Power ON and OFF Procedu...

Страница 18: ...rget mask by pulling it toward you Memo The target mask is held on by magnets 3 Take the desired target mask in your hand and attach it to the CM 3600A Make sure that it fits into the concave area aro...

Страница 19: ...uorescent calibration perform measurement under the condition that the UV cut off filter does not cover the xenon lamp UV light quantity is 99 9 Attaching the Zero Calibration Box Procedure 1 Pull the...

Страница 20: ...en performing white calibration the white calibration data for the white calibration plate being used is required The white calibration data for the white calibration plate included with the instrumen...

Страница 21: ...pen the cover of the transmittance chamber 2 Position the Transmittance Zero Calibration Plate in the transmittance chamber so that it completely blocks the illumination window Memo When using the opt...

Страница 22: ...transmittance chamber 2 Make sure that there is nothing specimen cell Transmittance Zero Calibration Plate etc between the illumination window and the receiving window in the transmittance chamber Mem...

Страница 23: ...around openings in the CM 3600A Failure to do so may result in fingers being trapped causing injury Reflectance measurement P 22 Transmittance measurement P 23 To measure the reflectance of a film or...

Страница 24: ...e reflected in the mirror inside the viewfinder cover can be observed When the viewfinder cover is opened the specimen will be illuminated for 60 seconds by a lamp to enable you to check the measuring...

Страница 25: ...it can be positioned with opposing surfaces parallel to both illumination and receiving windows Take care not to spill the liquid specimen on the CM 3600A Should there be any spillage immediately wipe...

Страница 26: ...emove dirt and dust from the target masks Do not touch the black coated surface of the target masks with fingers or wipe it with a cloth If the target masks get so dirty that dirt cannot be removed us...

Страница 27: ...enon lamp dead Illumination circuit failure Sensor failure Retry measurement or perform recalibration If the message persists contact the nearest Konica Minolta authorized service facility Not Calibra...

Страница 28: ...is incorrect Communication settings are incorrect Turn on the power of the instrument Check and correct the COM port and communication settings 15 Detected instrument disagrees with connected one nCon...

Страница 29: ...the operation manual of the software perform the operation correctly Transmittance measurement results seem strange Was specimen positioned correctly Position specimen correctly 23 Was 100 calibration...

Страница 30: ...28...

Страница 31: ...29 Explanations...

Страница 32: ...cimen in the transmittance chamber a The light transmitted by the specimen is received by the specimen measuring optical system and guided to the sensor b The diffused light in the integrating chamber...

Страница 33: ...hermore since the target masks are coated in black and measurement is influenced by the condition of this coated surface do not touch this surface with your hands scratch it or make it dirty Measureme...

Страница 34: ...sources SCI SCE SIMULTANEOUS MEASUREMENT Light sources and are located as illustrated at right Light source flashes first Memo Light source is a normal diffused type and enables the user to obtain SCI...

Страница 35: ...escent standard plate calibrated with a D65 light source are entered CALCULATING FLUORESCENT REFLECTANCE Based on the measured reflectances for the fluorescent standard plate obtained under the UV ful...

Страница 36: ...34 Dimensions 29 5 15 6 30 244 2 115 6 23 4 28 7 87 178 3 50 2 376 6 20 11 205 6 225 6 mm...

Страница 37: ...ble LAV 25 4 mm 30 mm MAV 8 mm 11 mm SAV 4 mm 7 mm Repeatability When white calibration plate is measured 30 times at 10 sec intervals after white calibration has been performed Spectral reflectance S...

Страница 38: ...36 MEMO...

Страница 39: ...RESULTING FROM THE MISUSE MISHANDLING UNAUTHORIZED MODIFICATION ETC OF THIS PRODUCT OR FOR ANY INDIRECT OR INCIDENTAL DAMAGES INCLUDING BUT NOT LIMITED TO LOSS OF BUSINESS PROFITS INTERRUPTION OF BUSI...

Страница 40: ...9222 A3UJ 12 BHEFDK 2011 2017 KONICA MINOLTA INC...

Отзывы: