K90100/K90190 Auto Melting Point Unit
Operation and Technical Manual
Page 26
Koehler Instrument Company, Inc.
1595 Sycamore Avenue • Bohemia, New York 11716-1796 • USA
Toll Free: 1-800-878-
9070 (US only) • Tel: +1 631 589 3800 • Fax: +1 631 589 3815
http://www.koehlerinstrument.com • e-mail: [email protected]
Petroleum Testing & Analysis Instrumentation • Custom Design & Manufacturing
NOTE:-
If ‘ENTER’ key is not pressed within 5 minutes, furnace will cool down to
room temperature.
The Lamp will be switched ON automatically as soon as the Furnace temperature
reaches 5°C below Set Temperature.
Display reads,
Instrument checks the light level & adjusts the illuminator
intensity automatically for the optimum sensitivity level of
Photo sensor.
In AUTO operation, if the sample is not placed properly / sample quantity is less /
sample preparation is not correct, a message is given accordingly.
Now, the furnace temperature rises as per user specified Rate.
For MANUAL operation, ‘MANUAL’ will appear in left
corner of the display instead of ‘AUTO’.
In AUTO operation, as soon as the sample starts melting, the Photo Sensor detects
the start and instrument automatically records it and disp
lays this temperature as ‘ST’.
Display reads,
Now further the Furnace temperature is raised as per the Heating Rate, when the
sample melts, instrument again automatically records it and display this temperature as
‘ET’. In
Display reads,
Switching Lamp
On…
T°C: 109.0
ST°C:____
AUTO
ET°C:____
T°C: 114.5
ST°C:114.5
AUTO
ET°C:____
T°C: 115.5
ST°C:114.5
AUTO
ET°C:115.5