K430X2 Dual Twin Foaming Characteristics Test Bath
Operation and Instruction Manual
K430X2-Manual
-
11-
diffuser located at the bottom of the Test
Cylinder.
8. Cold Bath.
For performing Sequence I and III
test procedures.
9. Hot Bath.
For performing Sequence II test
procedures.
4.2 Flowmeter Setup and Operation
1.
Connect the inlet on the flowmeter manifold
fitting to a source of clean dry air. Instrument
grade compressed air is recommended for this
application. Tygon or heavy wall rubber tubing
is suitable for connection to the manifold.
2.
Adjust the inlet air pressure to the regulator to
fifteen (15) psi for normal operation.
3.
Connect the outlets on the flowmeters to the
air tube assemblies using new ¼” tygon
tubing.
4.
Set the proper airflow by first consulting the
flowmeter calibration data in section 4.3. Find
the millimeter mark that will yield a flow of 94
ml/min.
5.
Adjust the valve on the flowmeter, so that the
ball float in the tube is centered on the
millimeter value specified in the data.
6.
Permit the air to flow for the duration specified
in the test procedure and disconnect the tygon
tubing from the flowmeter outlet when the
aeration period as elapsed.
4.3 Flowmeter Calibration Data
K43002-0-12 Air Flow Rate 94 mL/min
Float Material:
Sapphire
Float Density:
3.98 g/ml
Metered Fluid:
Air
Minimum Flow:
16.4 ml/min
Maximum Flow:
159.0 ml/min
Tolerance:
± 2% F.S
Scale Reading (mm)
Flowrate (ml/min)
65
159.0
60
139.0
55
121.2
50
104.7
45
90.8
40
77.8
35
66.0
30
56.0
25
46.5
20
38.3
15
30.5
10
23.0
5
16.4
4.4 Temperature Controller Operation
Figure 4. Temperature Controller
1. Process Temperature Display.
The
upper red LED display shows the process
temperature as read from the RTD probe.
2. Set Point Temperature Display.
The
lower green LED display shows the set
point temperature of the controller.
3. Programming Key.
Permits scrolling
through controller menu parameters. One
Level Forward
4. Down Key.
Used to decrease the set
point temperature and to decrease or
change
parameter
values
when
programming the temperature controller.
5. Up Key.
Used to increase the set point
temperature and to increase or change
parameter values when programming the
temperature controller.
6. Exit / Function Key.
This key is used to
exit or leave a level. One level backward
Содержание K430X2
Страница 2: ......
Страница 4: ......
Страница 5: ...K430X2 Dual Twin Foaming Characteristics Test Bath Operation and Instruction Manual K430X2 Manual 1 ...
Страница 6: ...K430X2 Dual Twin Foaming Characteristics Test Bath Operation and Instruction Manual K430X2 Manual 2 ...
Страница 23: ...K430X2 Dual Twin Foaming Characteristics Test Bath Operation and Instruction Manual K430X2 Manual 19 Notes ...
Страница 24: ...K430X2 Dual Twin Foaming Characteristics Test Bath Operation and Instruction Manual K430X2 Manual 20 Notes ...