K430X2 Dual Twin Foaming Characteristics Test Bath 

Operation and Instruction Manual 

 

K430X2-Manual

    

 

-

13-

 

5  Operation 

 

5.1  Bath 

 

Fill  each  bath  with  the  appropriate  heat transfer 

fluid  based  upon  the  testing  temperature.  Fill  the 
bath with the medium from the top through one of 
the seven port openings using a wide neck funnel 
with a 500mL griffin beaker adding about ½ L at a 
time to 2" (5 cm) from the top of the bath to allow 
room for fluid expansion.  

 

NOTE:

  Do  not  attempt  moving  the 

instrument when filled  with fluid medium. Drain all 
fluid  before  moving  or  relocating  the  instrument. 
See section 7.2 of this document. Sudden starting 
and  stopping  could  result  in  hot  bath  fluid  being 
forcefully ejected from the bath. It could also result 
in overbalancing and tipping over the instrument.  
 

NOTE:

  Do  not  drain  the  bath  until  the  fluid  is  at 

ambient temperature. 
 
 

Testing 

Temperature 

Recommended  Fluid 

15 - 50°C* 

 
Distilled water will be suitable. 
 

50 - 95°C 

Koehler 

supplies 

highly 

refined  white  technical  oil 
(part  #  355-001-001)  that 
contains an oxidation inhibitor 
to  limit  clouding  at  higher 
temperatures.  
CAS# 8042-47-5  
Minimum  Flash  Point:  120°C 
(248°F) 
Fire Point: 202°C (395°F) 
 

95 - 150°C 

Koehler  supplies  a  clear 
silicone  heat  transfer  fluid 
(part # 355-001-002) with high 
oxidation  resistance  and  low 
volatility.  CAS  #    63148-62-9. 
Minimum  Flash  Point:  200°C 
(392°F).  
 

 

NOTE:

 Do not use bath fluids that will 

decompose or volatilize producing flammable 
vapors below the maximum instrument operating 
tempe25°C.  
 

NOTE:

 Coloration of the Mineral or Silicon Bath 

Oil is an indication that the Medium must be 

changed. See Draining instructions in section 7.3 
of this document. When a decision is made to 
discard this material, dispose of this material and 
its container to a hazardous of special waste 
collection point. Country, State, or local laws may 
impose additional regulatory requirements 
regarding disposal. Dispose in accordance to all 
applicable regulations.  
 

WARNING:

 Bath oils contaminated with 

light ends can pose a flammability hazard.  

 

5.2  Power 

 

1.  Turn on the main power switches to the unit.  
 

WARNING:

 

Do 

NOT

  turn  the  power  on 

unless  the  bath  is  filled  with  the  proper  medium; 
otherwise,  damage  may  occur  to  the  unit  and  the 
warranty would be void.

 

 
2.  The temperature controllers will illuminate and 

display the actual and set point temperatures. 

 
3.  Set  each  bath  temperature  to  desired  test 

temperatures  as  per  ASTM  D892  using  the 
operating instructions in section 4.4.  

 

5.3  Preparation of Glassware and Tubing  

 

1.  Clean  the  graduated  cylinders  as  per  ASTM 

D892  instructions  and  pour  the  test  samples 
into the cylinders to the recommended level as 
per the aforementioned test methods.  

 
2.  Clean and flush the air delivery tube assembly 

with diffuser stone as per ASTM D892. 

 
3.  Place  the  four  1000  ml  test  cylinders  into  the 

tops  &  lock  in  place  with  the  four  clamps 
provided on each top.  

 
4.  Proceed  with  the  test  and  record  the  test 

results in accordance to ASTM D892. 

 
 
 
 
 
 
 
 
 

Содержание K430X2

Страница 1: ...ohemia New York 11716 1796 USA Toll Free 1 800 878 9070 US only Tel 1 631 589 3800 Fax 1 631 589 3815 http www koehlerinstrument com e mail info koehlerinstrument com Petroleum Testing Analysis Instrumentation Custom Design Manufacturing K430X2 Dual Twin Foaming Characteristics Test Apparatus ...

Страница 2: ......

Страница 3: ...with the applicable standards set forth in this certification Specifications ASTM D892 IP 146 DIN 51566 FTM 791 3211 FTM 791 3213 NF T 60 129 This unit is tested before it leaves the factory to ensure total functionality and compliance to the above specifications and ASTM standards Test and inspection records are on file for verification Jesse Kelly Application Engineer Koehler Instrument Company ...

Страница 4: ......

Страница 5: ...K430X2 Dual Twin Foaming Characteristics Test Bath Operation and Instruction Manual K430X2 Manual 1 ...

Страница 6: ...K430X2 Dual Twin Foaming Characteristics Test Bath Operation and Instruction Manual K430X2 Manual 2 ...

Страница 7: ...asic information on subsequent processing or recycling prior to disposal of the equipment http www dpa system dk en WEEE Products Private or professional use Responsibility for Registration and Annual Reporting Koehler will not sell directly to end users in the EU and so has no responsibility to register within each EU state and to make annual reports Koehler declares that this responsibility is b...

Страница 8: ...eparately collected WEEE Mercury containing components such as switches or backlighting lamps compact fluorescent lamps CFL Batteries Printed circuit boards if the surface of the printed circuit board is greater than 10 square centimeters about 4 sq in Toner cartridges liquid and pasty as well as color toner Chlorofluorocarbons CFC hydrochlorofluorocarbons HCFC or hydrofluorocarbons HFC hydrocarbo...

Страница 9: ... Flowmeter Setup and Operation 11 4 3 Flowmeter Calibration Data 11 4 4 Temperature Controller Operation 11 5 Operation 13 5 1 Bath 13 5 2 Power 13 5 3 Preparation of Glassware and Tubing 13 6 Safety Features 14 6 1 Over Temperature Protection 14 6 2 Over Power Protection 14 7 Maintenance 14 7 1 Routine Maintenance 14 7 2 Bath Cleaning 14 7 3 Replacement Parts 15 8 Wiring Diagrams 16 8 1 Unit Wiri...

Страница 10: ...K430X2 Dual Twin Foaming Characteristics Test Bath Operation and Instruction Manual K430X2 Manual 6 12 Warranty 17 13 Returned Goods Policy 18 Notes 19 ...

Страница 11: ...specialists who are experienced in both petroleum products testing and instrument service to better understand your requirements and provide you with the best solutions You can depend on Koehler for a full range of accurate and reliable instrumentation as well as support for your laboratory testing programs Please do not hesitate to contact us at any time with your inquiries about equipment tests ...

Страница 12: ... and Replacement Parts Any modification or alteration of this equipment from that of factory specifications is not recommended voids the manufacturer warranty product safety performance specifications and or certifications whether specified or implied and may result in personal injury and or property loss Replacement parts must be O E M exact replacement equipment Over Temperature Protection This ...

Страница 13: ...rom box to cart and from cart to bench Individuals must lift in accordance to proper technique See Figure below 7 Lift instrument from cart and place on bench 8 Ensure that all parts listed on the packing list are present Inspect the unit and all accessories for damage If any damage is found keep all packing materials and immediately report the damage to the carrier We will assist you with your cl...

Страница 14: ...1 Main Instrument Controls Figure 1 Main Instrument Descriptions 1 Power Switch This switch controls the power to motor and heater temperature controller of its corresponding side When the power switch is in the ON position the digital temperature controller and the stirrer are powered on 2 Temperature Controller The temperature controller regulates the bath temperature its corresponding bath Refe...

Страница 15: ...the duration specified in the test procedure and disconnect the tygon tubing from the flowmeter outlet when the aeration period as elapsed 4 3 Flowmeter Calibration Data K43002 0 12 Air Flow Rate 94 mL min Float Material Sapphire Float Density 3 98 g ml Metered Fluid Air Minimum Flow 16 4 ml min Maximum Flow 159 0 ml min Tolerance 2 F S Scale Reading mm Flowrate ml min 65 159 0 60 139 0 55 121 2 5...

Страница 16: ...is a value observed in the upper red LED display add it to the calculated difference obtained in the previous step This is the offset value c Press the Program Key The lower green display will flash Use the up or down keys to adjust to the new calibration offset value on the upper red display calculated in the previous step When the value has been entered the controller will automatically store th...

Страница 17: ...s that will decompose or volatilize producing flammable vapors below the maximum instrument operating temperature 25 C NOTE Coloration of the Mineral or Silicon Bath Oil is an indication that the Medium must be changed See Draining instructions in section 7 3 of this document When a decision is made to discard this material dispose of this material and its container to a hazardous of special waste...

Страница 18: ...e servicing to avoid exposure to high voltages and or temperatures which may result in personal injury or death If you have any questions about maintaining your equipment then please do not hesitate to contact the Koehler technical service department 7 1 Routine Maintenance The K430X2 Dual Twin Foaming Test Bath requires little routine maintenance to provide many years of continuous service Howeve...

Страница 19: ...0Hz 7A 278 001 002 Fuse 1A Slo Blo 5 X 20 mm 332 001 003 12 x 18 Borosilicate Bath Jar K23425 03000 18 Safety Shield K23700 03014A Motor 230V 50 60HZ 1 15HP K430 0 13 Air Outlet Elbow Tube K430 0 8 Rubber Stopper K430 3A Air Inlet Delivery Tube without diffuser K43002 0 12 Low Flowmeter K43002 0 13 Test Tube Rack K43002 0 15 Locking Post Assembly K43012 1000ml Cylinder w Ring K43002 0 9 Hot Bath O...

Страница 20: ...K430X2 Dual Twin Foaming Characteristics Test Bath Operation and Instruction Manual K430X2 Manual 16 8 Wiring Diagrams 8 1 Unit Wiring ...

Страница 21: ... Dual Twin Foaming Characteristics Test Bath should not require service Any service problem can be quickly resolved by contacting Koehler s technical service department either by letter phone fax or email In order to assure the fastest possible service please provide us with the following information Model Number Serial Number Date of Shipment 11 Storage This laboratory test instrument is equipped...

Страница 22: ...er the item s to be returned and the reason for the return You will be issued a Returned Authorization RA number which must be prominently displayed on the shipping container when you return the material to our plant Shipping containers without an RA number prominently displayed with will be returned to the sender Goods must be returned freight prepaid Returns will be subject to a restocking charg...

Страница 23: ...K430X2 Dual Twin Foaming Characteristics Test Bath Operation and Instruction Manual K430X2 Manual 19 Notes ...

Страница 24: ...K430X2 Dual Twin Foaming Characteristics Test Bath Operation and Instruction Manual K430X2 Manual 20 Notes ...

Отзывы: