![Koehler K39400 Скачать руководство пользователя страница 17](http://html1.mh-extra.com/html/koehler/k39400/k39400_operation-and-instruction-manual_1984934017.webp)
K394XX Water Separability Tester
Operation and Instruction Manual
K394XX-Manual
-
13-
7 Maintenance
WARNING:
Disconnect power to the unit
before servicing to avoid exposure to high
voltages and/or temperatures which may result
in personal injury or death. If you have any
questions about maintaining your equipment,
please do not hesitate to contact the Koehler
technical service department.
NOTE:
The bearings on the linear rail
are maintenance-free, lubrication-free. Do not
attempt to apply any lubricants to rail or
bearings. If sticking or slipping occurs, contact
the technical service department.
7.1
Routine Maintenance
The Koehler Water Separability Instrument
requires little routine maintenance to provide
many years of continuous service. However,
over the course of time, some instrument parts
may need to be replaced. A drain is also
provided in the back of the unit to empty the
bath of medium.
7.2
Bath Cleaning
•
To clean the instrument’s exterior, which
includes all painted surfaces and glass,
either a solution of soap and water or
laboratory grade detergent may be used.
•
Apply cleaner to clean wipe or cloth, not to
the instrument directly. Wipe surface clean.
•
Do Not
clean bath exterior with organic
chemicals such as Acetone, Toluene,
Hexane, etc.
•
For more difficult cleaning of finished
surfaces, a dilute solution or isopropanol in
water may be used.
•
It is not recommended that more aggressive
solvents be used on painted surfaces since
paint color will tarnish or be stripped from
the instrument.
•
Glass surfaces may be cleaned using a
more aggressive solvent such as acetone, if
necessary.
SHOCK AND BURN HAZARD:
Only
clean inside the bath when equipment is de-
energized and unplugged from the mains power
supply. Allow adequate time for heating coils to
completely cool before cleaning.
7.3
Replacement Parts
When ordering replacement part(s), please
provide the model number, serial number, and
product shipment date of your equipment so that
we can ensure you will receive the proper
replacement part(s).
Part Number
Description
050-001-028
Switch, Single Pole, 15A
050-001-044
Switch,
SPST, 15A
120/220V Rocker
091-032-001
Relay, Solid St, 4-
32VDC, 20A
265-122-002
RTD Temp Probe, 0.25 x
3’’, 2 Wire
271-015-004
Circuit Breaker, 2 Pole,
15A
275-103-027
Temp Cntrl, 100-240V,
W/RS232 BRD
332-002-018
Graduated Cylinder,
100mL
K39400-03054
Heater 1250W, 115V
K39400-03055
Heater 1250W, 220V
K23700-03013A
Motor, Modified, 115V
50/60Hz
K23700-03014A
Motor, Modified, 230V
50/60Hz
K39400-03025
Stirrer Shaft
282-014-016
Line cord, 115V
282-014-017
Line cord, 230V
240-230-004
Transformer 220 to 120V
Содержание K39400
Страница 2: ......
Страница 4: ......
Страница 19: ...K394XX Water Separability Tester Operation and Instruction Manual K394XX Manual 15...
Страница 20: ...K394XX Water Separability Tester Operation and Instruction Manual K394XX Manual 16...
Страница 21: ...K394XX Water Separability Tester Operation and Instruction Manual K394XX Manual 17 8 2 K39496 Unit Wiring...
Страница 22: ...K394XX Water Separability Tester Operation and Instruction Manual K394XX Manual 18...
Страница 23: ...K394XX Water Separability Tester Operation and Instruction Manual K394XX Manual 19...
Страница 25: ...K394XX Water Separability Tester Operation and Instruction Manual K394XX Manual 21 Notes...
Страница 26: ...K394XX Water Separability Tester Operation and Instruction Manual K394XX Manual 22 Notes...