K394XX Water Separability Tester 

Operation and Instruction Manual 

 
 

K394XX-Manual

    

 

-

8-

 

 

2.  Set Point Temperature Display.

  The 

lower green LED  display shows the set 
point temperature of the controller. 
 

3.  Programming Key.

  Permits scrolling 

through controller menu parameters. 
One Level Forward 
 

4.  Down Key.

  Used to decrease the set 

point temperature and to decrease or 
change parameter 

values 

when 

programming the temperature controller.  
 

5.  Up Key.

  Used to increase the set point 

temperature and to increase or change 
parameter values  when programming 
the temperature controller. 
 

6.  Exit / Function Key.

 This key is used to 

exit or leave a level. One level backward  
 

IMPORTANT NOTE:

 

The digital 

temperature controller for the unit 
comes pre-programmed from the 
Koehler factory. Please do NOT attempt 
to re-program the digital temperature 
controller as this will void the product 
warranty. If assistance is required, 
please do not hesitate  to contact the 
Koehler technical service department. 

 

Setting the Temperature

. Set the desired 

operating temperature by adjusting the set 
point with the up and down keys. The set 
point will be displayed in the lower green Set 
Point 

LED display and the actual 

temperature will be displayed in the upper 
red  Process  LED  display. Please allow the 
instrument 

to fully equilibrate before 

proceeding with any testing. 

 

Temperature Calibration.

  This routine allows 

the digital temperature controller to be 
calibrated to a certified thermometer. 

 

a.  Use a certified calibrated measuring 

device to acquire the temperature. 
Calculate the difference between the 
measuring device and the Process 
value displayed on the controller. 

b.  Press the program key two times until 

PCt

 is displayed in the lower green LED 

display. Press the DOWN key. CAL will 
display on the lower green display. If 
there is a value observed in the upper 

red LED display, add it to the calculated 
difference obtained in the previous step. 
This is the offset value. 

 

c.  Press the Program Key. The lower 

green display will flash. Use the up or 
down keys to adjust to the new 
calibration offset value on the upper red 
display calculated in the previous step. 
When the value has been entered, the 
controller will automatically store the 
value. The lower green display will stop 
flashing. If further adjustments are 
necessary, press the Program Key 
again. Resume regular operations by 
pressing the Exit / Function key two 
times. Verify if the new calibration is 
correct by observing the upper red 
display and comparing the value with 
the calibrated reference device.  

 
 

Auto Tune.

  This routine allows the digital 

temperature control to learn the heating 
parameters needed for any particular set 
point temperature. This operation should be 
done when installing a new unit, after 
replacing or changing the bath medium type, 
or utilizing a  different temperature set point 
20% different from the previously used set 
point temperature. 

 

a.  Set the operating temperature to the 

desired setting. 

 
b.  Press the up and down arrow buttons 

simultaneously for about 5 seconds. 
When Auto Tune is active, the lower 
green LED display will blink 

TUNE

. Auto 

Tune will automatically toggle off when 
the set point temperature is reached. 
Auto tune can be terminated by pressing 
the up & down buttons simultaneously 
again.  

 
 
 
 
 
 
 
 
 
 
 

Содержание K39400

Страница 1: ...venue Bohemia New York 11716 1796 USA Toll Free 1 800 878 9070 US only Tel 1 631 589 3800 Fax 1 631 589 3815 http www koehlerinstrument com e mail info koehlerinstrument com Petroleum Testing Analysis...

Страница 2: ......

Страница 3: ...andards set forth in this certification Specifications ASTM D1401 ASTM D6074 ASTM D6158 ISO 6614 DIN 51599 FTM 791 3201 NF T 60 125 This unit is tested before it leaves the factory to ensure total fun...

Страница 4: ......

Страница 5: ...tement is given cataloging the items that require separation from the equipment along with basic information on subsequent processing or recycling prior to disposal of the equipment http www dpa syste...

Страница 6: ...minimum the following substances preparations and components have to be removed from any separately collected WEEE Mercury containing components such as switches or backlighting lamps compact fluores...

Страница 7: ...packing 5 3 3 Setup 6 4 Descriptions 7 4 1 Description of Instrument Controls 7 4 2 Temperature Controller Operation 7 5 Operation 9 5 1 Bath 9 5 2 Power 9 5 4 Test Control Panel Operation 11 6 Safety...

Страница 8: ...ducts are backed by our staff of technically knowledgeable trained specialists who are experienced in both petroleum products testing and instrument service to better understand your requirements and...

Страница 9: ...Over temperature Protection This unit is equipped with over temperature Protection OTP circuitry to prevent overheating The unit will automatically interrupt power whether equipment malfunction or op...

Страница 10: ...red glass front panel jar pose a hazard to personnel Impacts to these during normal operation can shatter the glass components releasing the bath fluid posing thermal burn hazards and slipping hazards...

Страница 11: ...g Bath Stainless Steel Bath with clear tempered glass front viewing window for stabilizing the bath temperature to desired setting within 0 05 C 6 Cylinder Ports The test cylinder is placed into the h...

Страница 12: ...ture Calculate the difference between the measuring device and the Process value displayed on the controller b Press the program key two times until PCt is displayed in the lower green LED display Pre...

Страница 13: ...ium must be changed See Draining instructions in section 7 3 of this document When a decision is made to discard this material dispose of this material and its container to a hazardous of special wast...

Страница 14: ...e rail Otherwise shaft may get bent or damaged and will no longer be useful for operation 6 Lift the Stirrer shaft up then move the head assembly left or right until stirrer paddle is over the sample...

Страница 15: ...op button may be used to interrupt the 10 second audible alarm 11 At the end of the stirring period please follow the appropriate test method such as those listed in section 1 2 for determining the wa...

Страница 16: ...atus is equipped with Over temperature Protection OTP circuitry which prevents the unit from exceeding unsafe operating temperatures If the unit cannot maintain the set point temperature and begins to...

Страница 17: ...ior with organic chemicals such as Acetone Toluene Hexane etc For more difficult cleaning of finished surfaces a dilute solution or isopropanol in water may be used It is not recommended that more agg...

Страница 18: ...K394XX Water Separability Tester Operation and Instruction Manual K394XX Manual 14 8 Wiring Diagrams 8 1 K39400 Unit Wiring...

Страница 19: ...K394XX Water Separability Tester Operation and Instruction Manual K394XX Manual 15...

Страница 20: ...K394XX Water Separability Tester Operation and Instruction Manual K394XX Manual 16...

Страница 21: ...K394XX Water Separability Tester Operation and Instruction Manual K394XX Manual 17 8 2 K39496 Unit Wiring...

Страница 22: ...K394XX Water Separability Tester Operation and Instruction Manual K394XX Manual 18...

Страница 23: ...K394XX Water Separability Tester Operation and Instruction Manual K394XX Manual 19...

Страница 24: ...eturn of the product to the factory The sole responsibility of Koehler Instrument Company and the purchaser s exclusive remedy for any claim arising out of the purchase of any product is the repair or...

Страница 25: ...K394XX Water Separability Tester Operation and Instruction Manual K394XX Manual 21 Notes...

Страница 26: ...K394XX Water Separability Tester Operation and Instruction Manual K394XX Manual 22 Notes...

Отзывы: