K394XX Water Separability Tester 

Operation and Instruction Manual 

 
 

K394XX-Manual

    

 

-

9-

 

 

5  Operation  

 

5.1  Bath 

 

Fill the bath with the appropriate heat transfer 

fluid based upon the testing temperature. Fill the 
bath with the medium from the top through one 
of the seven port openings using a wide neck 
funnel with a 500mL griffin beaker adding about 
½ L at a time  to 2" (5 cm) from the top of the 
bath to allow room for fluid expansion. This will 
provide the proper depth for immersing the 
viscometers and allow for thermal expansion. 

 

NOTE:

  Do not attempt moving the bath 

when filled with fluid medium. Drain all fluid 
using the drain valve at the back of the 
instrument before moving or relocating. Sudden 
starting and stopping could result in hot bath 
fluid being forcefully ejected from the bath. It 
could also result in overbalancing and tipping 
over the instrument.  
 

NOTE:

 Do not drain the bath until the fluid 

is at ambient temperature. 
 
 

Testing 

Temperature 

Recommended  Fluid 

15 - 50°C* 

 
Distilled water is 

Mandatory

 

 

50 - 95°C 

Koehler supplies highly 
refined white technical oil 
(part # 355-001-001) that 
contains an oxidation inhibitor 
to limit clouding at higher 
temperatures.  
CAS# 8042-47-5  
Minimum Flash Point: 120°C 
(248°F) 
Fire Point: 202°C (395°F) 
 

 
 

NOTE:

 Do not use bath fluids that will 

decompose or volatilize producing flammable 
vapors below the maximum instrument operating 
tempe25°C.  
 

NOTE:

 Coloration of the Mineral Oil is an 

indication that the Medium must be changed. 

See Draining instructions in section 7.3 of this 
document. When a decision is made to discard 
this material, dispose of this material and its 
container to a hazardous of special waste 
collection point. Country, State, or local laws 
may impose additional regulatory requirements 
regarding disposal. Dispose in accordance to all 
applicable regulations.  
 

WARNING:

 Bath oils contaminated with 

light ends can pose a flammability hazard.  

 

5.2  Power 

 

Turn on the main power switch to the unit.  

 

WARNING:

 

Do 

NOT

  turn the power on 

unless the bath is filled with the proper medium; 
otherwise, damage may occur to the unit and 
the warranty would be void.

 

 

5.3  Starting a Test 

         

Before running a test, make sure the unit is 
level. Otherwise, the results will be affected. 
 

1. 

Install an appropriate ASTM thermometer or 
PT100 Temperature in the cover plate of the 
bath using the holder provided.

 

 

2. 

Verify Temperature controller calibration in 
accordance to an external temperature 
reader and set the desired operation 
temperature using the temperature controller 
(See Section 4.3 for full operational details).  

 

3. 

Prepare and place up to seven samples into 
the sample holders on the top front of the 
unit and lower until bottom of cylinder is in 
complete contact with the metal base.  

 

WARNING:

 

Do not test or run materials 

that boil or exhibit high vapor pressure at the 
bath test temperature.  

 

IMPORTANT:

 Allow the bath 

AND

 the sample 

enough time to reach the temperature before 
starting measurements. Allow 10-15 minutes 
for the bath to equilibrate. Do not add medium 
to the bath while a test is being performed. 
Additional medium will disrupt the temperature 

Содержание K39400

Страница 1: ...venue Bohemia New York 11716 1796 USA Toll Free 1 800 878 9070 US only Tel 1 631 589 3800 Fax 1 631 589 3815 http www koehlerinstrument com e mail info koehlerinstrument com Petroleum Testing Analysis...

Страница 2: ......

Страница 3: ...andards set forth in this certification Specifications ASTM D1401 ASTM D6074 ASTM D6158 ISO 6614 DIN 51599 FTM 791 3201 NF T 60 125 This unit is tested before it leaves the factory to ensure total fun...

Страница 4: ......

Страница 5: ...tement is given cataloging the items that require separation from the equipment along with basic information on subsequent processing or recycling prior to disposal of the equipment http www dpa syste...

Страница 6: ...minimum the following substances preparations and components have to be removed from any separately collected WEEE Mercury containing components such as switches or backlighting lamps compact fluores...

Страница 7: ...packing 5 3 3 Setup 6 4 Descriptions 7 4 1 Description of Instrument Controls 7 4 2 Temperature Controller Operation 7 5 Operation 9 5 1 Bath 9 5 2 Power 9 5 4 Test Control Panel Operation 11 6 Safety...

Страница 8: ...ducts are backed by our staff of technically knowledgeable trained specialists who are experienced in both petroleum products testing and instrument service to better understand your requirements and...

Страница 9: ...Over temperature Protection This unit is equipped with over temperature Protection OTP circuitry to prevent overheating The unit will automatically interrupt power whether equipment malfunction or op...

Страница 10: ...red glass front panel jar pose a hazard to personnel Impacts to these during normal operation can shatter the glass components releasing the bath fluid posing thermal burn hazards and slipping hazards...

Страница 11: ...g Bath Stainless Steel Bath with clear tempered glass front viewing window for stabilizing the bath temperature to desired setting within 0 05 C 6 Cylinder Ports The test cylinder is placed into the h...

Страница 12: ...ture Calculate the difference between the measuring device and the Process value displayed on the controller b Press the program key two times until PCt is displayed in the lower green LED display Pre...

Страница 13: ...ium must be changed See Draining instructions in section 7 3 of this document When a decision is made to discard this material dispose of this material and its container to a hazardous of special wast...

Страница 14: ...e rail Otherwise shaft may get bent or damaged and will no longer be useful for operation 6 Lift the Stirrer shaft up then move the head assembly left or right until stirrer paddle is over the sample...

Страница 15: ...op button may be used to interrupt the 10 second audible alarm 11 At the end of the stirring period please follow the appropriate test method such as those listed in section 1 2 for determining the wa...

Страница 16: ...atus is equipped with Over temperature Protection OTP circuitry which prevents the unit from exceeding unsafe operating temperatures If the unit cannot maintain the set point temperature and begins to...

Страница 17: ...ior with organic chemicals such as Acetone Toluene Hexane etc For more difficult cleaning of finished surfaces a dilute solution or isopropanol in water may be used It is not recommended that more agg...

Страница 18: ...K394XX Water Separability Tester Operation and Instruction Manual K394XX Manual 14 8 Wiring Diagrams 8 1 K39400 Unit Wiring...

Страница 19: ...K394XX Water Separability Tester Operation and Instruction Manual K394XX Manual 15...

Страница 20: ...K394XX Water Separability Tester Operation and Instruction Manual K394XX Manual 16...

Страница 21: ...K394XX Water Separability Tester Operation and Instruction Manual K394XX Manual 17 8 2 K39496 Unit Wiring...

Страница 22: ...K394XX Water Separability Tester Operation and Instruction Manual K394XX Manual 18...

Страница 23: ...K394XX Water Separability Tester Operation and Instruction Manual K394XX Manual 19...

Страница 24: ...eturn of the product to the factory The sole responsibility of Koehler Instrument Company and the purchaser s exclusive remedy for any claim arising out of the purchase of any product is the repair or...

Страница 25: ...K394XX Water Separability Tester Operation and Instruction Manual K394XX Manual 21 Notes...

Страница 26: ...K394XX Water Separability Tester Operation and Instruction Manual K394XX Manual 22 Notes...

Отзывы: