A-62051 May 2008
11
Adjusting the developer
temperature control
The Archive Processor’s developer temperature control is factory set to
a value of 88.0 to control the developer temperature at 37.8°C ±0.3°C
(100°F ±0.5°F). The display values and setpoint value are arbitrary
numbers and do not represent the temperature in degrees; they are
values relative to the developer temperature.
Reading the developer temperature
The developer temperature must be read on the dial thermometer.
1. Lift the processing section cover.
2. Locate the dial thermometer in the front right corner of the #1 rack.
3. Read the value on the thermometer.
If the developer temperature thermometer value is not 37.8°C
±0.3°C (100°F ±0.5°F), you must make an adjustment. Adjust the
developer temperature control setpoint value only if necessary.
Refer to the next section, “Changing the developer temperature
control setpoint value.”
4. Close the processing section cover.
Thermometer