8.4 Resetting the adjustment
You can reset the electronic analyzer so that it once again uses its full range
(0 - 2.0 m/s)
Rotate the flow potentiometer (1) clockwise as far as it will go (to the right-hand
stop).
Rotate the temperature potentiometer (2) counterclockwise as far as it will go
(to the left-hand stop).
Press and hold the SET button (5) until the flow LED stops flashing
The device is now reset to its initial state (0 - 2 m/s).
8.5 Switching point setting
After the device is adjusted to the specific liquid being measured (see above), you
can then set the switching point.
Adjust the flow rate.
Rotate the flow potentiometer (1) counterclockwise as far as it will go (to the
left-hand stop). The first LED in the LED strip display (3) starts to blink.
Any number of LEDs in the LED strip display (3) is illuminated constantly.
Rotate the flow potentiometer (1) clockwise to the desired switching point.
The switching point selected is indicated by the blinking LED.
Display 3
X
X X X X
X X
X
X X X
X X
X
X X X
X X X X X
X
X X X
explanation
output
flow > setpoint
Flow relays
activated
Flow LED lights
flow = setpoint
Flow relays
activated
Flow LED lights
flow < setpoint
Flow relays
deactivated
Flow LED off
blinks
Flow speed
LED indicator blinks
Содержание KAL-0312 EX
Страница 1: ...Operating Instructions for Flow sensor model KAL Ex with Electronic Analyzer Model KAL E Ex...
Страница 17: ...KAL EX KAL EX K05 0322 page 17 16 EC type examination certificate...
Страница 18: ...KAL EX page 18 KAL EX K05 0322...
Страница 19: ...KAL EX KAL EX K05 0322 page 19 17 IECEx certificate...
Страница 20: ...KAL EX page 20 KAL EX K05 0322...
Страница 21: ...KAL EX KAL EX K05 0322 page 21...
Страница 22: ...KAL EX page 22 KAL EX K05 0322...
Страница 23: ...KAL EX KAL EX K05 0322 page 23...