8.6  Temperature compensation 

If the length of the cable connecting the sensor and the electronic analyzer is greater than 10 
meters or if increased accuracy in temperature measurement is desired, we recommend that 
device be compensated as described below. 

 
 

Caution! Be sure to switch off the power supply 

 

  Connect the entire length of the cable to be used to terminals 6 and 8. 

Instead of the sensor, connect a 100 ohm ±1% resistor (included in the delivery). 

  Set the temperature potentiometer (2) to 80° C (rotate to right-hand stop) Switch on the 

power source and place the electronic analyzer in operation. 

  Briefly press the SET button (5). 

The temperature LED and the flow LED will start to blink. After 10 seconds, they will go 
out.  

  If the system is not functioning properly, the flow LED (6) will blink continuously. In this 

case, press the SET button (5) once again. The device then switches back to measuring 
mode. Before continuing, inspect the mechanical and electrical installation. Then follow the 
“Temperature compensation” instructions once again. 

  If the temperature LED blinks faster, then the error could be caused because the sensor is 

still connected instead of the 100 ohm resistor. 

  Finally, remove the 100 ohm resistor and reconnect the sensor. 

 
 
 

9. Fault alarms 

In alarm mode the current relay stops functioning.  
This occurs under the following conditions: 
 

  Actual value < Setpoint value 

  Short circuit (at sensor) 

  Line break 

  Power failure 

 
 
 

10. Maintenance 

This  device  is  maintenance-free.  Once  a  month,  the  sensor  tip  should  be  checked  for  the 
presence of mineral deposits (lime, etc.) and cleaned if necessary. 

Содержание KAL-0312 EX

Страница 1: ...Operating Instructions for Flow sensor model KAL Ex with Electronic Analyzer Model KAL E Ex...

Страница 2: ...nt 8 8 3 Expanding the trend indicator 8 8 4 Resetting the adjustment 9 8 5 Switching point setting 9 8 6 Temperature compensation 10 9 Fault alarms 10 10 Maintenance 10 11 Brief operating instruction...

Страница 3: ...devices are only to be used maintained and serviced by persons familiar with these operating instructions and in accordance with local regulations applying to Health Safety and prevention of accident...

Страница 4: ...by a series of constantly illuminated LEDs on the LED strip display The currently set switching point is indicated on the LED strip display by the individually blinking LED Sensors These sensors are...

Страница 5: ...he manufacturer s specifications may invalidate its warranty Therefore any resulting damage is not the responsibility of the manufacturer The user assumes all risk for such usage 5 Operating Principle...

Страница 6: ...tend at least 5 mm better 5 mm into the pipe see below 6 3 Mounting position The sensors can be mounted in any position provided that the piping is completely filled with the liquid to be monitored Th...

Страница 7: ...ice Connect the sensor cable to terminals 6 and 8 of the electronic analyzer Connect the supply line power supply to terminals 15 and 16 Caution The relays are shown in the rest positions The intrinsi...

Страница 8: ...point pen The flow LED 6 will blink for approximately 30 seconds After this time has expired the flow LED 6 will stop blinking If the system is not functioning properly the flow LED 6 will blink cont...

Страница 9: ...ic liquid being measured see above you can then set the switching point Adjust the flow rate Rotate the flow potentiometer 1 counterclockwise as far as it will go to the left hand stop The first LED i...

Страница 10: ...to blink After 10 seconds they will go out If the system is not functioning properly the flow LED 6 will blink continuously In this case press the SET button 5 once again The device then switches bac...

Страница 11: ...g range Turn to the right hand stop 2 Potentiometer for setting the temperature switching point SET button 5 for adjustment expanding measuring range 6 Flow LED illuminates when flow is greater than t...

Страница 12: ...ints compensation at potentiometer Switching function relay is operated with flow terminal 9 10 closed LED illuminated relay is operated when temperature is exceeded LED illuminated terminal 12 13 clo...

Страница 13: ...1210 G 1 2 KAL 1315 KAL 1415 KAL 1115 KAL 1215 G 3 4 KAL 1320 KAL 1420 KAL 1120 KAL 1220 G 1 KAL 1125 KAL 1225 G 11 4 KAL 1132 KAL 1232 G 11 2 KAL 1140 KAL 1240 M 12 x 1 KAL 0312 1 4 NPT KAL 5308 KAL...

Страница 14: ...s flats A B C D A B C D E M 12x1 SW 19 23 43 G 1 4 SW 27 10 50 81 G 1 4 SW 19 26 43 G 1 2 SW 27 10 50 81 G 1 2 SW 27 43 58 G 3 4 SW 32 15 52 82 5 G 3 4 SW 32 43 58 G 1 SW 39 15 56 85 1 4 NPT SW 19 26...

Страница 15: ...s Part 0 General Requirements EN 60079 11 2012 Explosive atmospheres Part 11 Equipment protection by intrinsic safety i Ebenfalls mit folgenden Europ ischen und nationalen Normen und technischen Vorsc...

Страница 16: ...8 done at Hofheim on April 11th 2018 Name des Unterzeichners Name of signatory Name des Unterzeichners Name of signatory H Peters Gesch ftsf hrer CEO Manfred Wenzel Prokurist authorized signatory Unte...

Страница 17: ...KAL EX KAL EX K05 0322 page 17 16 EC type examination certificate...

Страница 18: ...KAL EX page 18 KAL EX K05 0322...

Страница 19: ...KAL EX KAL EX K05 0322 page 19 17 IECEx certificate...

Страница 20: ...KAL EX page 20 KAL EX K05 0322...

Страница 21: ...KAL EX KAL EX K05 0322 page 21...

Страница 22: ...KAL EX page 22 KAL EX K05 0322...

Страница 23: ...KAL EX KAL EX K05 0322 page 23...

Отзывы: